THE BATTERY STATUS INDICATOR (BSI)
BSI Color
Battery Charger Mode
Solution
Off
Battery pack not sensed in well or charger
Place battery pack in well and/or check all power
base not connected to power supply.
connections.
Green
Battery pack fully charged.
Battery pack is ready for use.
Amber
Battery pack in well charging, but not
Continue charging until green light appears.
ready.
Amber flashing
Battery pack in well deep discharging.
Continue charging until green light appears.
Red flashing
Bad battery pack or charger base.
Replace battery pack. If problem persists with new
(battery pack in well)
battery pack, charger base has malfunctioned. Call
for service.
Contact your local dealer or Respironics if frequent errors occur.
13.0 TROUBLESHOOTING GUIDE
At times, you may receive an error code (E01, E02, etc.) or warning symbol in the display window. The following is a
guide to help you correct a system problem.
Display
Window
Indication
Solution
E00
No user set-up information identified.
Install a charged battery pack into the battery compartment.
Press
F1
and
F2
at the same time, hold for one second and
then release. This places the unit in the Set-up mode. Enter
settings specific to your institution, see section 12.0 for
Set-up mode instructions.
E01
Calibration measurement is unacceptable.
Press and release
F1
and
F2
at the same time. Ensure the
Bili
Cal
™ Tip is in place. Press the
TRIGGER
button to
restart calibration (---). If error continues try another
Bili
Cal
Tip. If error persists, remove the Bili
Cal
and clean
the optical measurement tip (under disposable tip) with
approved cleaning agent, repeat calibration. If error
continues replace measurement light bulb (see section 15.0
for replacement instructions) and repeat calibration. If error
E01 persists, call for service.
E02
Patient measurement is unacceptable.
Press and release
F1
and
F2
at the same time. Retest the
patient. If E02 occurs again, replace Bili
Cal
, repeat
calibration and retest patient. If E02 continues, call for service.
E03
Disposable Tip removed between the calibration and
Press and release
F1
and
F2
at the same time. Remove the
measurement scan or the Bili
Chek
is being used
Bili
Cal
. Replace with a new tip and recalibrate. If E03
in an environment with too much ambient light.
error is indicated even though the Disposable Tip has not
been removed, the ambient light level is too high. The
Bili
Chek
should not be used outdoors, too close to infra-red
radiant warmers, or in areas with high levels of sunlight.
Move to an area of lower light levels and press the
F1
and
F2
buttons at the same time.
E04
Too much movement during the measurement.
Press and release
F1
and
F2
and retest the patient.
Measurement is out of clinically proven range.
Obtain a Total Serum Bilirubin level (TSB).
E10
Light bulb is burned out.
Replace measurement light bulb, see section 15.0 for
replacement instructions. If error continues, call for service.
E99
System error is detected.
Call for service.
Low battery pack, there are approximately 10
Replace current battery pack with a fully charged battery
measurements left when the icon first appears.
from the charger base and place discharged battery pack into
well “B”for charging.
Discharged battery pack, no measurements
Replace current battery with a fully charged battery pack
can be taken.
from the charger base and place discharged battery pack into
well “B” for charging.
E05
HIGH
21