Connecting the humidifier bottle

Connect the oxygen outlet in panel and the oxygen inlet in top of humidifier with the 

oxygen connection tube.

Turning on the power 

Plug in power cord to on an electrical outlet.

Selecting the working mode 

Start  the  device  if  needs  oxygen  inhalation,  adjust  at  the  flow-rate  advised  by 

physician.

Select  the  compressed  oxygen  output  mode  if  needs  the  compressed  oxygen  output 

function.

Oxygen inhalation

Connect an end of oxygen tube to the outlet of humidifier bottel,fix the tube on use

.

Cautions: 

  Don't  adjust  the  oxygen  inhalation  time  or  the  flow-rate  and  Check  the 

oxygen  tubes  are  connected  well.  Don't  pull  the  tubes  in  strength,  avoiding  the  tubes 
falling down. oxygen therapy should listen to the physician’s advice.

's 

ears, insert the other end into uses' nose for inhalation

Timing

Set the time according to the timer function.

During  treatment,  there  will  be  alarm  if  the  external  power  off,  the  user  should  take 

down  the  oxygen  tube  and  check  the  power  cord,  using  it  continuously  after 
troubleshooting.  Press  On/Off  button  to  use  it  continuously  after  the  external  power 
regaining.

End of use

Press  the  On/Off  button  to  stop  the  device  working  after  finished  using  and  turn  off 

the power cord. Take down the power line if you don’t often use it.

5.4

 

 Operating Panel and LCD Screen

Start/Stop

Plus Key 

Minus Key

Timer

 Key

/Mode

Oxy Concentrtor

Oxygen

0

1

2

3

4

Total time 

00007

 hour

Flow

L/min

It is composed of buttons, indicator light and LCD screen, etc.

The LCD screen  display the standby interface and the operation interface.

S t a n d b y  i n t e r f a c e :

  S h o w  t h e 

working  mode,  accumulated  service 
time.

Flow regulator

0xygen outlet

Respiratory Medical Solutions

START/STOP

Oxygen

Concentration

0

1

2

3

4

This time 00:00 h/min
Timer off   - -:- -  h/min

Flow

L/min

Timer

 Key

/Mode

Start

/

Stop:

Start or stop operation

In working  state,press this button to enter into time mode

In the standby state,switch between the oxygen supply mode(CFM) and

compressed air outlet mode

(

AO5W) by

 

pressing

 

this

 

button

Plus/Minus Key

In timing mode,use this button to increase or reduce the time

Respiratory Medical Solutions

 

13

12

Содержание AO3 Series

Страница 1: ...piratory equipment www rms care com EC REP Part NO 233010002 Rev 3 0 3 20 15 Manufacture Shenyang RMS Medical Tech Co Ltd Address No 8 1 Gaoge Road Hunnan New District Shenyang China Tel 0086 024 23782895 8001 Fax 0086 024 23782895 8008 KINGSMEAD SERVICE LIMITED 145 157 St John Street London EC1V 4PY UK TEL 044 20 7193 9159 ...

Страница 2: ...e Warranty Instructions Section 7 Troubleshooting Guide Troubleshooting Alarms Appendix A EMC Information 04 04 05 06 06 07 07 07 07 07 08 08 09 10 11 11 11 11 13 14 15 15 15 16 16 17 18 Drip Proof Equipment IP21 NO smoking Serial Number Indicating its conformity with the medical devices directive 93 42 EEC The 0917 is the number of notified body Proper disposal of equipment according to local law...

Страница 3: ...ction 1 Special Notes Please read all instructions before using Important information is highlighted by these terms 1 1 Warnings Never use aerosol sprays or flammable anesthetics near the equipment If the equipment has a damaged cord or plug if it is not working properly if it has been dropped or damaged or if it has been submersed in water do not use and call a qualified technician for examinatio...

Страница 4: ...ency so it may cause interference if using the portable communication equipments near the device 1 4 Intended Use Medical oxygen concentrator Model AO3 AO3W AO5 AO5W used for separating oxygen from air EUT not intended to be used for emergency treatment and patient who has serious carbon monoxide poisoning cannot use this device Section 2 Packaging And Handling 2 1 Unpacking Check for any obvious ...

Страница 5: ...test Oxygen Concentration Testing Circuit Breaker Power Switch Power Socket Filter Device Tag LCD Screen Control Buttons Caste Handle Warning Tag Oxygen Outlet Inlet Filter Window Compressed Air Output Humidifier Bottle Section 3 Features 3 1 Parts and Functions Flow Regulating Button Filter Oxygen Outlet Control Buttons Respiratory Medical Solutions 09 08 ...

Страница 6: ...backpressure is 0 and the rated flow is 3L min giving a 7kpa backpressure can make the flow being 2 8 L min If the power network isn t stable the power voltage fluctuate over 10 please use it after installing a voltage stabilizer Make sure the environment nearby dry ventilated no dust and any other caustic gas no strong electromagnetic interference This device can be used normally below the height...

Страница 7: ...en tube and check the power cord using it continuously after troubleshooting Press On Off button to use it continuously after the external power regaining End of use Press the On Off button to stop the device working after finished using and turn off the power cord Take down the power line if you don t often use it 5 4 Operating Panel and LCD Screen Start Stop Plus Key Minus Key Timer Key Mode Oxy...

Страница 8: ...e The damage caused by human error fault operation konck moving liquid infiltration etc The damage by natural disasters or accidents Other accessories not listed for warranty Oxygen Concentration 0 1 2 3 4 This time 00 00 h min Timer off h min Flow L min 5 5 Operating Method Operation interface Display working mode alarm information oxygen concentration oxygen flow present service time timing time...

Страница 9: ...ndicator flashing Continuous Alarm Concentrator not operating Press start stop the alarm stops Call service provider indicator flashing Continuous Alarm start stop the alarm stops Call service provider Concentrator not operating The red low pressure indicator flashing Continuous Alarm Press start stop the alarm stops Call service provider Concentrator not operating The red hign pressure indicator ...

Страница 10: ...3 3 Vrms 150 kHz to 80 MHZ outside ISM bandsa 3 V m 80 MHz to 2 5 GHz 3V 3V m Portable and mobile RF communications equipment should be used no closer to any part of the XXXXXXX including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Guidance and manufacturer s declaration electromagnetic immunity This device is intended...

Отзывы: