ATF System with C410:V4B Controller
User Guide
79
d.
Set InputPage3.P2_P2_Capture = 1 until the OutputPage4.P2_Cal_Enable bit is
received.
i.
Wait for the
“RequestedDataPage”
to be equal to 4; since cycling through
output data pages is recommended.
ii.
Examine OutputPage4.P2_
P2_Capture. When true…
e.
Reset InputPage3.P2_P2_Capture = 0; the maximum capture bit is momentarily
asserted.
6)
Assert accept bit:
a.
While “SendingDataPage” is still equal to 3…
b.
Set InputPage3.P2_Accept = 1 until the OutputPage4.P2_Accept bit is received.
i.
Wait for the
“RequestedDataPage”
to be equal to 4; since cycling through
output data pages is recommended.
ii.
Examine
OutputPage4.P2_Accept. When true…
c.
Reset InputPage3.P2_Accept = 0.
d.
Controller performs the linear scaling calculation and resets the controller’s
internal calibration enable bit.
7)
Clean up: After all calibration activities are complete, be sure to perform the following:
a.
Reset parameters that were set during the calibration process.
b.
Set “SendingDataPage” = 0 to prepare for next ATF run commands.