XCell
TM
ATF System with C410:V4B Controller
User Guide
50
On User logoff, the controller continues to operate normally. A user must log back on to make
changes to setting.
6.4
Startup
Assembly of the
XCell™ ATF
System is described in the following sections. This section
provides a startup guide for an
XCell™ ATF
System connected to a bioreactor and ready for
use, with the following general conditions:
•
An
XCell™ ATF
6 System is used in this example.
•
An
XCell™ ATF
System flow of 12L/min is required.
•
Bioreactor pressure is 0.0psi
•
∆
height between pump midpoint and vessel liquid level is 0cm
•
Regulated air pressure set to 35psi
•
Vacuum source connected
6.4.1 Following LOGON and entry of BATCH Information, go to the SET UP screen.
6.4.2 Go to the BASIC SET UP screen and enter the following field values.
Field
VALUE
Controller Set UP
Pump Model No
Select
XCell™ ATF
6
Control Mode
Select FLOW
Slope Function Enable
Do not enable
Bioreactor pressure
0
∆
height (cm)
0
Alarm Delay(min)
1
Alarm Set Points
HiHi Alarm Set Points
7 psi
LoLo Alarm Set Point
-7 psi
Alarm Delay (x10 msec)
50 for both
Pump Parameters
P-Pressure Offset (psi)
0.5
E-Pressure Offset (psi)
-0.5
P-Delay (%)
70
E-Delay (%)
70
P-Overtime (%)
120
E-Overtime (%)
120
P-PV Step Size (%)
3
E-PV Step Size (%)
3
6.4.3 Press Accept Changes.
H.
LOGOFF.