Operating Instructions 

 

 

 

 
5.0 Mounting the accelerometer 

 

Important! 
 
Sensor mounting: 
The accelerometer should generate signals for the movement and acceleration 
of the feeder, which are fed back to the regulator circuit of the control unit. 
Therefore it is very important that no other extraneous vibration signals are 
picked up by the sensor 

 
 

 

 

 
6.0 Calibrating the sensor input 

Important! 

 

The  sensor  constants  deviating  from  the  indicated  sensor  voltage  (mV/g)  must  be 
set once on the measuring unit! 

 

A  calibration  certificate  is  provided  with  each  sensor  supplied.  This  certificate  shows  the 

exact value of the sensor output voltage, for example 95 mV/g. This value must be entered in the Menu: 
Service with Parameter “Cal”. 
The sensor input is factory-set to 100 mV/g. It is possible to set deviations up to +/- 20 %. 
 
 
7.0 Saving the measured values 

All four values (acceleration, speed, deflection / amplitude, frequency) are saved. 
Press the arrow key to select the no. of the set to start the storage with. 
 
In Menu: Storage, Parameter: Record 

 0-39 

 

 
 
 

8.0 Read measurements 

This menu serves to control the data sets stored. The individual data sets may be riewed and cleared (if 
necessary) by pressing the arrow keys. 
 
In Menu: Records, Parameter: Record 

 0-39 

 
 
 
 
 
 
 

Sensor

Swing Direction

Mounting Block

Moving Part

Fixed Part

Wrong

Correct

Sensor

Sensor

Содержание REOVIB SMW 4000

Страница 1: ...6774_EN_ANL_rev00 DOC 23 12 2014 REOVIB For Vibratory Feeder Systems SWM 4000 Handheld measuring instrument for vibratory feeder ...

Страница 2: ...ure 6 5 0 Mounting the accelerometer 7 6 0 Calibrating the sensor input 7 7 0 Saving the measured values 7 8 0 Read measurements 7 9 0 Saving your own parameter 8 10 0 Function only measure or enable and disable Parameter Menues 8 11 0 Factory setting 8 12 0 Readout the saved data sets into a PC 8 13 0 Delivery contents 9 14 0 Spare Parts Single components 9 15 0 Dimensions 10 ...

Страница 3: ...nal safety of operators and also for the protection of the described product and connected equipment Warning Hazardous Voltage Failure to observe can kill cause serious injury or damage Isolate from mains before installation or dismantling work as well as for fuse changes or post installation modifications Observe the prescribed accident prevention and safety rules for the specific application Bef...

Страница 4: ...low versatile operation Optimization of the design display and internal battery means that the unit is faster more reliable and yet is smaller and more convenient to use The REOVIB SMW4000 instrument is supplied with a comprehensive list of accessories consisting of USB cable and charger traveler protective cover 100mV g accelerometer sensor mounting bracket and sturdy carry case 2 0 Technical Dat...

Страница 5: ...level or return to Operation mode 5 6 7 Select menu Open Close Menu Select parameters 1 Indicates that the Sensor is not connected or Sensor fault blinking Acceleration g Speed cm s Deflection mm Frequency Hz Acceleration g Speed cm s Deflection mm Frequency Hz I 0 P P P 1 Info Records Service Storage black background Selected menu parameters 2 Acceleration 5 00 g Speed 7 5 cm s Deflection 0 23 mm...

Страница 6: ... Acceleration sensor calibration 8 600 mV g 4 3 Language Select the menu language wählen 4 4 Multi Line Changing the menu structure Multi Line Sin gle Line 1 0 4 5 Back Light Backlight off on 1 0 4 6 Factory Settings Restoring factory settings Execute 4 7 Code Menus enabling or disabling see section 9 0 and 10 0 4 8 User Params Load customers parameter settings Execute 4 9 STOP Auto Off Auto Off 1...

Страница 7: ...or supplied This certificate shows the exact value of the sensor output voltage for example 95 mV g This value must be entered in the Menu Service with Parameter Cal The sensor input is factory set to 100 mV g It is possible to set deviations up to 20 7 0 Saving the measured values All four values acceleration speed deflection amplitude frequency are saved Press the arrow key to select the no of t...

Страница 8: ...Parameter Code 117 Lock Menues 1 Enabling parameter Menus In Menu Service Parameter Code 117 Lock Menues 0 11 0 Factory setting With this parameter you can restore the factory settings In Menu Service Parameter Factory Settings 12 0 Readout the saved data sets into a PC At this unit it is possible to load the saved data sets via the USB interface into a PC For the coupling you need Microsoft EXCEL...

Страница 9: ...000 Sturdy carry case Sensor mounting bracket Accelerometer 100mv g incl mounting magnet 14 0 Spare Parts Single components Designation Order Number Measuring instrument SWM 4000 with USB charger power Supply traveler incl USB da ta charging cable 677411 USB charger power Supply traveler incl USB data charging cable 090611 Silicone protective cover for SWM 4000 090612 Sturdy carry case 090613 Sens...

Страница 10: ...REOVIB SWM 4000 Operating Instructions 10 15 0 Dimensions ...

Страница 11: ...REOVIB SWM 4000 Operating Instructions 11 ...

Страница 12: ...reo ag in Italy REO ITALIA S r l Via Treponti 29 I 25086 Rezzato BS Tel 39 030 279 3883 Fax 39 030 279 0600 E Mail info reoitalia it Internet www reoitalia it Poland REO CROMA Sp zo o ul Pozaryskiego 28 bud 20 PL 04 703 Warszawa Tel 48 0 22 812 3066 Fax 48 0 22 815 6906 E Mail croma croma com pl Internet www croma com pl Spain REO ESPAÑA 2002 S A C Manuel Ventura i Campeny 21B local 9 E 08339 Vila...

Отзывы: