TONiC RSLM20/RELM20 Functional Safety installation guide and safety manual           Original instructions

System calibration

Calibration is an essential operation that completes readhead set-up, with the optimum incremental  

and reference mark signal settings stored in the readhead’s non-volatile memory. 

Before system calibration

 

X

Clean the scale and readhead optical window (contamination around the reference mark may result in 
reference mark dephasing).

 

X

If re-installing the readhead, restore factory defaults.

 

X

Maximise the signal strength along full axis of travel.

NOTE:

  Calibration (CAL) routine should be carried out at slow speed (<100 mm/s or less than the 

readhead maximum speed, whichever is slowest).

Step 1 – Incremental signal calibration

 

X

Ensure Automatic Gain Control (AGC) is switched off – Calibration (CAL) LED on readhead is not 
illuminated – before beginning calibration. 

 

X

Press and release the CAL button on the end of the interface using a 2 mm Allen key or similar tool as 

shown. 

NOTE:

 Activating the CAL switch only requires 2.5 N force. Applying excess force may permanently 

damage the switch.

Set-up LED

CAL button

 

X

The CAL LED will now periodically single-flash to indicate that it is in incremental signal calibration 

mode.

 

X

Move the readhead along the axis, ensuring you do not pass the selected reference mark, until the CAL 
LED starts double-flashing, indicating the incremental signal is now calibrated and the new settings are 

stored in the readhead memory. 

 

X

The system is now ready for reference mark phasing.

 

X

For systems without reference mark, go to ‘Calibration routine – manual exit’.

 

X

If the system does not automatically enter the reference mark phasing stage (no double-flashing 

of the CAL LED) the calibration of the incremental signals has failed. After ensuring failure is not due 

to overspeed (>100 mm/s or exceeding the readhead maximum speed), exit the calibration routine, 

restore factory defaults and check the readhead installation and system cleanliness before repeating the 
calibration routine.

Step 2 – Reference mark phasing

 

X

Move the readhead back and forth over the selected reference mark until the CAL LED stops flashing 

and remains off. The reference mark is now phased. 

 

X

The system automatically exits the CAL routine and is ready for operation. 

 

X

For the safety function to be valid, customer commissioning tests must be performed:  

see ‘Commissioning test’ on page 3

.

 

X

If the  CAL LED continues double-flashing after passing the chosen reference mark many times, it is not 

detecting the reference mark. 

 

-

Ensure that the correct readhead configuration is being used. Readheads can either output all 
reference marks or only output a reference mark where a reference selector magnet is fitted.

 

-

Check that the reference selector magnet is fitted in the correct location relative to readhead 

orientation 

(see ‘TONiC FS T301x readhead installation drawing’ on page 6)

.

NOTE:

 The LED on the interface will flash blank when the reference mark is detected (>100 mm/s only). 

 

It indicates the presence of a reference mark, not the phasing status.

Calibration routine – manual exit

 

X

To exit the calibration routine at any stage press the CAL button. The CAL LED on the readhead will then 
stop flashing. 

 

X

For the safety function to be valid, customer commissioning tests must be performed:  

see ‘Commissioning test’ on page 3

.

Restoring factory defaults

When re-installing the system, or in the case of continued calibration failure, factory defaults should be 
restored.

To restore factory defaults

 

X

Switch system off.

 

X

Press and hold the CAL button whilst switching the system on. The CAL LED on the readhead will flash 
several times, indicating that the factory defaults have been restored.

 

X

Release CAL button.

 

X

Chec

‘Readhead mounting and alignment’ on page 14

 and re-calibrate the system.

NOTE:

 System must be re-calibrated after restoring factory defaults. 

Switching AGC on or off

AGC can be switched on or off via the interface.

 

X

Press and hold the CAL button on the interface for >3 seconds to switch AGC on or off. 

 

The CAL LED on the readhead will be illuminated when AGC is active.

NOTE: 

The system must be calibrated before switching AGC on.

Ti interface

DOP interface

CAL LED

Settings stored

Single-flashing

None, restore factory defaults and re-calibrate

Double-flashing

Incremental only

Off (auto-complete)

Incremental and reference mark

Содержание TONiC FS T301 Series

Страница 1: ...Installation guide M 6688 9046 02 B TONiC FS T301x RSLM20 RELM20 high accuracy linear encoder system Functional Safety installation guide and safety manual...

Страница 2: ...Cable connection 12 System connection Ti interface 12 System connection DOP interface 13 Readhead mounting and alignment 14 System calibration 15 Diagnostic LEDs 16 Output signals 17 Speed 18 Electri...

Страница 3: ...case the user will be required to correct the interference at their own expense NOTE This unit was tested with shielded cables on the peripheral devices Shielded cables must be used with the unit to...

Страница 4: ...r own assessment of the full system to determine its safety capability Correct use includes X X Operating the TONiC FS system within the limits defined in this document X X Installing the system as de...

Страница 5: ...gnals falls within the limits defined in Evaluation unit monitoring Evaluation unit monitoring To achieve full system integrity the evaluation unit must continuously monitor the analogue outputs and i...

Страница 6: ...AHF79 The products comply with EU directives and comply with the following technical standards 2014 30 EU Electromagnetic compatibility EMC 2011 65 EU On the restriction of the use of certain hazardou...

Страница 7: ...e CH3 CH2 5CH3 Propan 2 ol CH3CHOHCH3 Scale and readhead Acetone CH3COCH3 Methylated Spirits Chlorinated Solvents Readhead only Acetone CH3COCH3 Methylated Spirits Chlorinated Solvents Scale only Mini...

Страница 8: ...2 5 0 2 5 35 23 11 5 6 min 2 mounting holes M2 5 through counterbored 3 2 3 deep both sides Set up LED CAL AGC LED 0 6 Pitch tol 1 Rideheight 2 1 0 15 Alternative mounting face 2 mounting holes M2 5...

Страница 9: ...anual Original instructions Measuring lengths Dimensions and tolerances in mm 20 ML L 10 without limits Measuring length ML L 25 with dual 10 mm limits Optical centreline incremental and reference mar...

Страница 10: ...from scale ends Double sided adhesive tape is included with all scale lengths When scale is to be mounted horizontally on a vertical surface position the dowels so that the datum edge is supported Ad...

Страница 11: ...remaining backing liner Area for RGG 2 epoxy A 9531 0342 6 Support dowels 7 Remove excess epoxy or 5 Locate scale against dowels or ledge and rotate down onto substrate Press down with firm finger pre...

Страница 12: ...concentrators on the readhead The design of the bracket should be sufficiently stiff so that it is able to tolerate such force without distorting Following the clamping instructions on the scale insta...

Страница 13: ...of travel as indicated by the readhead and interface set up LEDs readhead Green DOP interface ideally Blue Purple CALIBRATION Press and release the CAL button on the interface The CAL LED on the readh...

Страница 14: ...ween the readhead and the interface X X The customer is responsible for verifying the product function where the readhead to interface cable has been re terminated including the installation of extens...

Страница 15: ...aking care not to touch the pins plug the connector into the socket in the interface ensuring correct orientation as shown 3 Refit the cover plate ensuring the cable ferrule is located in the recess o...

Страница 16: ...arance will contribute to mechanical safe position Green Orange Red Readhead set up LED status Yaw 0 0 4 Pitch 0 1 Roll 0 0 5 Rideheight 2 1 0 15 mm DOP interface set up LED status Blue Green Orange R...

Страница 17: ...d back and forth over the selected reference mark until the CAL LED stops flashing and remains off The reference mark is now phased X X The system automatically exits the CAL routine and is ready for...

Страница 18: ...evel 50 to 70 No Red Poor set up signal may be too low for reliable operation signal level 50 No Purple blank flashing Over signal system in error Yes Blue blank flashing Over speed system in error Ye...

Страница 19: ...14 Shield Inner shield Not connected Outer shield Case Function Output type Signal Colour Power 5 V Brown 0 V White Incremental signals Analogue Cosine V1 Red Blue Sine V2 Yellow Green Reference mark...

Страница 20: ...40 2 70 1 35 0 540 0 270 0 135 0 054 0 027 25 10 10 8 10 3 24 1 62 0 810 0 324 0 162 0 081 0 032 0 016 20 10 10 6 75 2 70 1 35 0 675 0 270 0 135 0 068 0 027 0 013 12 10 9 4 50 1 80 0 900 0 450 0 180 0...

Страница 21: ...0 V 5 V Readhead IMPORTANT The outer shield must be connected to the machine earth Field Ground The inner shield must be connected to 0 V at receiving electronics only Care should be taken to ensure t...

Страница 22: ...Select standard or wide reference at time of ordering to match the requirements of the controller being used 20 m 45 Output specifications Analogue output signals Incremental 2 channels V1 and V2 diff...

Страница 23: ...rating 0 C to 70 C Sealing readhead IP40 Ti interface IP20 DOP interface IP30 Acceleration readhead Operating 500 m s 3 axes Shock system Non operating 500 m s 6 ms sine 3 axes Vibration system Operat...

Страница 24: ...20 Selectable reference marks every 200 mm RELM20 Midpoint of scale length RELE20 20 mm from end of scale Phasing Auto phased by readhead calibration routine Mounting Self adhesive backing tape Repeat...

Страница 25: ...e enable disable Push and release 3 seconds AGC enable disable Push and hold during power Off On cycle Restore factory defaults Refer to readhead LED functionality chart for set up LED indications 48...

Страница 26: ...T 2019 Renishaw plc All rights reserved Renishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renish...

Отзывы: