Renishaw RGH40 Скачать руководство пользователя страница 6

Dimensions and tolerances in mm

RGS40-G glass scale installation drawing 

(adhesive mounting)

RGH40 RGS40-G

 

installation guide

4

Support dowel alignment method

Ledge alignment method

Reference mark alignment line

Reference mark  

sensor position

Optical centreline

Bolted Q limit 
magnet  

(A-9550-0048)

Scale graduations

Reference mark actuator 

(A-9531-0250)

*

M3 x 0.5  
x >6.5

2.9

Ø4

>10

3

18

8.5

4.05

17.5

25.5

12.5

>27

5

ML + 47

ML + 44

L

ML

14.5

(L-25) 

(n-1)

4.9

2.9

10 
12

RGG-2, 2-part epoxy usually coincident 
with user selected scale datum

Double-sided adhesive tape

*

Bolted P limit 
magnet 

(A-9550-0047)

0.5

0.1/100

F

F

0.05

F = axis guideway

*

Double-sided adhesive tape is included with all scale lengths

Adhesive datum clamp

Limit magnets for 
ledge mounting 

(A-9531-0251)

Scale length   L (mm)

130

180

230

280

310

400

510

760

1010

Measuring length   ML (mm)

120

170

220

270

300

390

500

750

1000

Recommended number of support dowels, n

3

3

3

3

3

3

3

4

5

Содержание RGH40

Страница 1: ...Installation guide M 9550 9023 01 A RGH40 RGS40 G linear encoder system...

Страница 2: ...ss scale installation drawing clip clamp mounting 6 Clip clamp mounting installation 7 Reference mark actuator installation 8 Limit switch installation 8 Readhead mounting and alignment 8 Reference ma...

Страница 3: ...tion of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense NOTE This unit was tested wi...

Страница 4: ...cal encoder scale that provides good immunity against contaminants such as dust fingerprints and light oils However in harsh environments such as machine tool applications protection should be provide...

Страница 5: ...es optical and reference mark centreline Roll tol 1 Q limit magnet A 9531 0251 Mount dimple up for Q Optional bolted limit magnet P A 9550 0047 Q A 9550 0048 Mounting ledge or top surface of bracket f...

Страница 6: ...6 5 2 9 4 10 3 18 8 5 4 05 17 5 25 5 12 5 27 5 ML 47 ML 44 L ML 14 5 L 25 n 1 4 9 2 9 10 12 RGG 2 2 part epoxy usually coincident with user selected scale datum Double sided adhesive tape Bolted P lim...

Страница 7: ...to cure fully then clean scale using Renishaw scle wipes A 9523 4040 or a clean dry lint free cloth Adhesive mounting installation 2 Mark area for epoxy on adhesive tape 1 RGH40 RGS40 G installation...

Страница 8: ...x 6 5 14 5 M3 x 0 5 x 6 5 23 5 0 1 34 4 10 3 18 8 5 4 05 17 5 25 5 12 5 27 5 5 5 Datum clamp Mounting clip 4 9 2 9 Dimensions and tolerances in mm RGS40 G glass scale installation drawing clip clamp m...

Страница 9: ...ean dry lint free cloth M 9550 0057 gap setting shim 4 Ensure scale can move so free to expand contract 1 2 mm 1 2 mm M 9550 0057 gap setting shim M 9550 0061 datum pad RGH40 RGS40 G installation guid...

Страница 10: ...able conformance to the installation tolerances allow adjustment to the rideheight of the readhead and be sufficiently stiff to prevent deflection or vibration of the readhead during operation For eas...

Страница 11: ...way D type plug L 12 way circular V 12 way circular coupling W 16 way in line connector X Power 5 V Brown 4 2 2 A Brown link 5 12 12 M 0 V White 12 10 10 B White link 13 11 11 N Incremental signals V1...

Страница 12: ...ers must ensure they comply with the lowest recommended counter input frequency Encoder velocity m s Resolution m x 4 safety factor Function Signal Colour 15 way D type plug D 16 way in line connector...

Страница 13: ...d from each other If the inner and outer shields are connected together this will cause a short between 0 V and earth which could cause electrical noise issues Electrical connections Grounding and shi...

Страница 14: ...P Resolution S A B Z Inverse signal not shown for clarity Synchronised pulse Z duration as resolution S Repeatability of position uni directional maintained within 10 C from temperature at time of pha...

Страница 15: ...ealing IP50 Acceleration Operating 500 m s2 3 axes Shock Non operating 1000 m s2 6 ms sine 3 axes Vibration Operating 100 m s2 max 55 Hz to 2000 Hz 3 axes Mass Readhead 50 g Cable 38 g m Cable 12 core...

Страница 16: ...our main website at www renishaw com contact M 9550 9023 01 RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countrie...

Отзывы: