DCO-0599
Effective Date: 14-DEC-2018
LBL-6025 Rev B
Page 4 of 8
9)
Scrub the Shaver Blade/RF Probe Cavity using the Shaver Blade
Cavity cleaning brush for a minimum of 1 minute.
10)
Scrub the Motor Shaft using the Motor Shaft cleaning brush for a
minimum of 1 minute.
11)
Scrub the Fluid Channel using the Fluid Channel cleaning brush
for a minimum of 1 minute.
12)
Ensure that all interior surfaces, crevices, lumens, cavities, and
holes of the handpiece are scrubbed thoroughly to remove any
visible debris.
13)
Scrub the exterior surface of the handpiece with a gauze pad
soaked in the cleaning solution for a minimum of 1 minute.
14)
Remove the handpiece from the cleaning bath and rinse it
thoroughly, flushing the interior lumens and cavities, for a
minimum of 2 minutes under running lukewarm tap water 72° -
109°F (22° - 43°C).
Cleaning Verification
15)
After cleaning, inspect the handpiece under normal lighting to
ensure that all visible debris has been removed.
16)
If not visibly clean, repeat cleaning and re-inspect.
17)
For difficult-to-view areas, 3% hydrogen peroxide may be applied
(bubbling is evidence of the presence of blood).
Note
: Rinse the handpiece thoroughly under running lukewarm
tap water 72° - 109°F (22° - 43°C) following any hydrogen
peroxide testing.
18)
Shake the handpiece with the shaver blade cavity end pointing
downward to remove any water.
19)
Dry the handpiece with a lint-free towel.
Note
: Compressed air or nitrogen may be used to remove any
additional moisture from the handpiece.
10.2.
STERILIZATION
INSTRUCTIONS
Clean and sterilize the handpiece prior to the first use and every use
thereafter.
Caution
Never sterilize the handpiece that has not been cleaned.
The success of the sterilization process depends on the
previous state of cleaning.
Caution
Always install the soaking cap prior to cleaning or
sterilization. If moisture reaches the connection pins of
the cable connector, it could damage the connector.
Caution
The handpiece can be sterilized by using the following
methods only.
Sterilizing the Handpiece:
1)
Clean the handpiece according to the “Cleaning Instructions”
section of this manual.
2)
Attach the soaking cap to the cable connector if it is not already
attached.
Caution
Do not cross-sterilize the handpiece. Alternating
between different methods of sterilization may cause
undue wear on the handpiece. Product longevity is
extended by the consistent use of a single sterilization
method.
Caution
Remove the handpiece from the sterilizer immediately
after sterilization, or product damage may result.
Caution
Never immerse the handpiece in a liquid to help it cool
after sterilization, or product damage may result. If
necessary, cool the handpiece by wrapping it in a
sterilized damp sponge or cloth.
3)
Prepare the handpiece for sterilization by either:
a) Wrapping the handpiece in Kimberly-Clark Kimguard KC400
Sterilization Wrap or equivalent
b) Following the steps below if the optional Sterilization Tray is to
be used.
4)
Place the handpiece into the optional Sterilization Tray as shown
in the figure below. Ensure that the handpiece is snapped into
the rubber holders.