background image

DCO-0599 

Effective Date: 14-DEC-2018 

 

LBL-6025 Rev B 

Page  5 of 8 

 

 

 

5)

 

Coil the cable up adjacent to the handpiece.  Ensure that the 

soaking cap is placed on the cable connector 

6)

 

Place the lid onto the sterilization tray. 

 

7)

 

Wrap the sterilization tray in Kimberly-Clark Kimguard KC400 

Sterilization Wrap or equivalent.  

8)

 

Sterilize the handpiece according to the parameters listed for one 

of the following methods: 

STERRAD 100S Method: 

1)

 

Clean the handpiece according to the “Cleaning Instructions” 

section of this manual. 

2)

 

Ensure the soaking cap is installed. 

3)

 

Run STERRAD 100S Short Cycle Sterilization System per the 

manufacturer’s instructions. 

STERRAD 100NX Method: 

1)

 

Clean the handpiece according to the “Cleaning Instructions” 

section of this manual. 

2)

 

Ensure the soaking cap is installed. 

3)

 

Run STERRAD 100NX STANDARD Cycle Sterilization System 

per the manufacturer’s instructions. 

STERRAD NX Method: 

1)

 

Clean the handpiece according to the “Cleaning Instructions” 

section of this manual. 

2)

 

Ensure the soaking cap is installed. 

3)

 

Run STERRAD NX Standard Cycle Sterilization System per the 

manufacturer’s instructions. 

STERIS V-PRO Method: 

1)

 

Clean the handpiece according to the “Cleaning Instructions” 

section of this manual. 

2)

 

Ensure the soaking cap is installed. 

3)

 

Run the Lumen Cycle on any of the following STERIS V-PRO 

Sterilization Systems per the manufacturer’s instructions:            

V-PRO 1, V-PRO PLUS, V-PRO MAX, V-PRO MAX 2. 

11.  STORAGE 

The handpiece and accessories should be stored either in their 

shipping box or in a sterilization tray.  In either case, ensure that the 

Tricera Handpiece is immobile to prevent any damage. 

12.  PERIODIC MAINTENANCE SCHEDULE 

To ensure the safety and longevity of your handpiece, observe the 

following maintenance procedures. 

Before every surgical procedure: 

1)

 

Assemble the entire system, including the controller, footswitch, 

handpiece, and shaver blade/RF probe. 

Caution 

 

To ensure proper heat dissipation and keep the joints 

clear, do not run the handpiece without proper suction or 

irrigation.  Immerse the blade in a sterile irrigant and 

activate suction while testing the handpiece. 

2)

 

Run the handpiece through its various functions, modes, and 

settings to ensure it responds properly to commands.  Ensure that 

all parts move freely and that there are no loose or missing 

components.  Note any unusual sounds or vibrations, and note 

operating speeds. 

3)

 

Refer any calibration or operating difficulties to your Relign sales 

representative or Relign Customer Service. 

Every 12 Months: 

Check the leakage current using a true RMS digital multimeter and 

safety analyzer.  Consult the analyzer’s instructions for details. 

13.  LIMITED WARRANTY 

Relign warrants to the original purchaser of Tricera Handpiece that the 

Tricera Handpiece will be free of defects in material and workmanship 

for a period of one year after the date of purchase when used as 

intended under normal operating conditions and in conformance with its 

instructions for use and maintenance instructions.  Any warranty 

concerning software and/or hardware upgrades shall expire in the 

original one-year warranty.  The sole obligation of Relign and its 

affiliates under this warranty shall be limited to the replacement of the 

Tricera Handpiece or equivalent models, at no charge to the original 

purchaser and at the option of Relign, during the warranty period but 

only if examination reveals to the satisfaction of Relign that the Tricera 

Handpiece contains a defect in material or workmanship.

 

Relign and its 

affiliates will not be liable for any incidental, special or consequential 

loss, damage, or expense resulting, directly or indirectly, from the use 

of its product regardless of whether Relign or its affiliates have been 

advised of the possibility of such loss, damage or expense. 
This warranty is made in lieu of all other warranties, expressed, implied 

or statutory, including, but not limited to the implied warranties of 

merchantability and fitness for a particular purpose and all other 

obligations and liabilities on the part of Relign.  Relign neither assumes 

nor authorizes any other person to assume for it any other liability in 

connection with the sale of the Tricera Handpiece.  This warranty shall 

not apply to the Tricera Handpiece or any other part thereof that has 

Содержание R-10002

Страница 1: ...asis of blood vessels through its shaver blade and probe Examples of uses of the product include resection of torn knee cartilage subacromial decompression and resection of synovial tissue in other joints Contraindications The electrosurgical probe should not be used in procedures where a nonconductive irrigant is used or with patients having cardiac pacemakers or other electronic implants 4 CONTR...

Страница 2: ...ssory or device be sure to follow the instructions provided with the accessory or device 7 HOW SUPPLIED Warning The Tricera Handpiece is supplied non sterile It must be cleaned and sterilized before the first use Prior to each subsequent use it must be cleaned and sterilized The shipping box contains 1 ea Relign Tricera Handpiece 1 ea Shaver Blade RF Probe Cavity Brush Gordon Brush PN 721012 or eq...

Страница 3: ...ece from the other system components a Remove the shaver blade RF probe from the handpiece b Dispose of all used shaver blades RF probes according to hospital regulations c Disconnect the handpiece from the Relign Controller 2 Attach the soaking cap to the cable connector Caution Always install the soaking cap prior to cleaning or sterilization If moisture reaches the connection pins of the cable ...

Страница 4: ...ny additional moisture from the handpiece 10 2 STERILIZATION INSTRUCTIONS Clean and sterilize the handpiece prior to the first use and every use thereafter Caution Never sterilize the handpiece that has not been cleaned The success of the sterilization process depends on the previous state of cleaning Caution Always install the soaking cap prior to cleaning or sterilization If moisture reaches the...

Страница 5: ...ade in a sterile irrigant and activate suction while testing the handpiece 2 Run the handpiece through its various functions modes and settings to ensure it responds properly to commands Ensure that all parts move freely and that there are no loose or missing components Note any unusual sounds or vibrations and note operating speeds 3 Refer any calibration or operating difficulties to your Relign ...

Страница 6: ... term original purchaser as used in this warranty means that person or organization and its employees if applicable to whom the Tricera Handpiece was sold by Relign This warranty may not be assigned or transferred in any manner In order to make a warranty claim the original customer must obtain a Return Materials Authorization Number RMA Number and return instructions from Relign s Customer Servic...

Страница 7: ...era Handpiece 1 Shaver Blade RF Probe Cavity Brush Gordon Brush PN 721012 or equivalent 2 Fluid Channel Brush Key Surgical PN AT 12 197 or equivalent 3 Motor Shaft Brush Gordon Brush PN 499NAD ST or equivalent Figure A3 Cleaning Brushes 1 Sterilization Tray 2 Sterilization Tray Lid 3 Rubber Holders Figure A4 Optional Sterilization Tray if available 1 2 3 1 Shaver Blade Cavity 2 Motor shaft 3 Fluid...

Страница 8: ...The shaver blade does not turn The shaver blade is not inserted properly Remove the blade and reinsert it according to the Relign Tricera System instructions for directions on use of the Tricera Handpiece as part of the Relign Tricera System The handpiece will not run The handpiece is not properly connected to the controller Remove and reconnect the handpiece cable to the controller according to t...

Отзывы: