background image

16

    

   

17

15180-101 Rev. F

15180-101 Rev. F

Instrument Setup

 (continued)

Instrument Settings

 (continued)

Practice Screen

The Practice Screen allows you to enter a business name or other message that will automatically print 

with the patient data.

Creating a Message

1.  On the Practice ID screen, press the blue UP, 

DOWN, RIGHT, or LEFT icons to highlight the 

character that you want to appear in the text 

box. 

2.  Select the ENTER icon.  The selected 

character will appear in the text box.   

Refer to Figure SU-14.

3.  Repeat Step 1 and 2 until you finish your 

message.

4.  When you are finished, highlight DONE and 

select the ENTER icon to save your message 

and return to the Setup Screen.  

5.  This Practice ID will now print out with the 

patient data when you select YES under 

Display ID# in the Setup Screen.

Editing Characters in a Message

1.  On the Practice ID screen, use the left and right arrow icons in the main selection area, to 

highlight the character 

in front of

 the character you want to change in the text box.

2.  Use the UP, DOWN, RIGHT, or LEFT icons to highlight a replacement character from the grid, 

then select the ENTER icon.  The character will be inserted 

behind

 the highlighted letter.  Refer 

to Figure SU-13.

3.  Repeat steps 1 and 2 for each character that needs to be changed.

Deleting Characters in a Message

1.  On the Practice Screen, use the LEFT, RIGHT, UP, or DOWN icons to highlight the character 

in 

front of

 the character that needs to be deleted.  Select the CLEAR icon, located above the DONE 

icon.

2.  To erase the entire message and start from the beginning, use the UP, DOWN, RIGHT, or LEFT 

buttons to highlight the CLEAR icon, and select it.  The entire message will be deleted.

Figure SU-14  ID Screen

Text Box

Содержание Lensometer LensChek Plus

Страница 1: ...LensChek Plus Pro Digital Lensometer User s Guide...

Страница 2: ...reserves the right to make changes in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 13485 Certified Reichert produc...

Страница 3: ...20 Lens Placement 20 Single Lens 21 Framed Lens 22 Multifocal Lens 23 Progressive Lens 24 Contact Lens 27 UV Measurement LensChek Pro 28 PD Measurement LensChek Pro 28 Marking Lenses 29 Lenses With As...

Страница 4: ...AND TESTING MUST BE CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING THE INTERNAL CIRCUITRY OF THE INSTRUMENT CONTAINS ELECTROSTATIC DISCHARGE SENSITIVE DEVICES ESDS THAT MAY BE SENS...

Страница 5: ...MAGE TO THE DISPLAY SEE MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTION CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS REGARDING EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVIC...

Страница 6: ...lity Directive 2014 30 EU and Low Voltage Directive 2014 35 EU Consult Instructions for Use Indicates that important operating and maintenance instructions are included in this User s Guide Fragile Co...

Страница 7: ...eye care technicians should operate this instrument All parts of this ME system are suitable for use within the patient environment Note Words in all capital letters reference menu choices Note Words...

Страница 8: ...of the box 9 Remove the accessories from the Accessory Box This box contains the following accessories Note The Printer Paper and the Contact Lens Stand can be placed in the Storage Compartment for s...

Страница 9: ...ht or left Nose Pad should be positioned in the middle of the frames to obtain a PD measurement LensChek Pro only Lens Stand Where the lens is placed in order to take a reading Power Switch Use this s...

Страница 10: ...d Marking Pens Use these pens to place ink marks on the lens Lens Stand Install this cover on the Nosepiece to measure eyeglass lenses Contact Lens Cover Install this cover on the Nosepiece to measure...

Страница 11: ...to the LensChek Plus Pro is within the listed voltage range Connect the power cord to the LensChek Plus Pro and then connect the other end to the power source Set the Power switch to the ON position p...

Страница 12: ...me of normal measurement they correspond to the icons shown below Clear Icon Clears the measurement values The data stored in memory is also cleared by selecting this icon S R L R L Icon Selects singl...

Страница 13: ...e SU 08 Control Button Icons for Setup screen Up Icon Select this icon to move up the screen or to a different page Left Icon Select this icon to move left on the screen Right Icon Select this icon to...

Страница 14: ...ent screen is reflected here Step Selects step to display measurement value 0 25 0 12 0 01 Auto Prog Sets auto detection of progressive lens On Off Lens Selects lens to be measured Glasses H CL S CL G...

Страница 15: ...Baud Rate Select the baud rate 115200 38400 9600 Prog Graph Sets display of ADD value and assessment graph on progressive lens measurement screen On Off Graph Print Sets printing of ADD value and ass...

Страница 16: ...he unit goes into sleep mode the power light will blink This is to indicate it is in sleep mode Pressing any button will wake up the unit and turn it back on Tone Enables an audible tone to sound when...

Страница 17: ...lay ID in the Setup Screen Editing Characters in a Message 1 On the Practice ID screen use the left and right arrow icons in the main selection area to highlight the character in front of the characte...

Страница 18: ...bles the LensChek Plus Pro printer If you press the Printer Icon on the LCD Screen the patient data goes to the LensChek Plus Pro printer RS232C Enables the RS232C port If you press the TRANSFER icon...

Страница 19: ...s the PLUS or MINUS buttons to change the date values 4 To change the Time Format press the UP or DOWN arrows to reach the Time Format field Press the RIGHT or LEFT buttons to select a time format Ref...

Страница 20: ...For instructions on using the Marking Pens refer to the Marking Lenses section of this manual Note For instructions on placing and measuring contact lenses refer to the Contact Lens section of this m...

Страница 21: ...te the lens to fit the axis direction Note The Alignment Mark represents the optical center of the lensmeter and the Cross Cursor represents the optical center of the lens 4 Press the Memory Add butto...

Страница 22: ...center of the lens 5 Press the Memory Add button to store the Sphere Cylinder Axis and Prism measurements Refer to Figure OP 07 Note If Auto Memory in the Setup screen is set to On then the lens will...

Страница 23: ...of near point by moving the lens so that the near point near sight segment comes to the center of the lens stand Note A measurement can be taken even if the messages Align ment OK and Marking OK are n...

Страница 24: ...ssive lens measurement screen If not the measurement screen remains in single focus lens mode Select the Progressive Icon to switch to the progressive mea surement screen Refer to Figure OP 15 Note If...

Страница 25: ...e cross cursor will change from green to blue and the gray box will turn blue indicating the values are saved Refer to Figure OP 18 3 Take a measurement of near point by moving the lens slowly so that...

Страница 26: ...ning the near and far points To take a measurement of far point manually perform the same steps described in the Measuring Progressive Lenses section of this manual 1 In the Setup Menu set Prog Graph...

Страница 27: ...lens Re fer to Figure OP 23 6 The message Alignment OK will appear on the screen when the lens is aligned properly Refer to Figure OP 24 Note The Alignment Mark represents the optical center of the l...

Страница 28: ...Pull out the lens table so that the bottom of the lens touches the lens table and the corresponding Nose Pad is aligned in the center of the glasses Refer to Figure OP 27 4 When measuring the lens the...

Страница 29: ...fingers Do not lower the pens without first placing a lens on the Lens Stand Mark your lenses gently Do not mark a lens multiple times in the same location Lenses With Astigmatism For a lens with asti...

Страница 30: ...2 Move the lens so that the prism values displayed on the screen match with the ones on the pre scription Refer to Figure OP 35 P Prism Value B Base Direction Lens Offset 1 Select on from Prism mm on...

Страница 31: ...D is set Maximum number of characters 44 characters 22 characters X 2 lines The unit of the prism value is different ac cording to the setting Measurement value of right lens When the ADD value and as...

Страница 32: ...e soft cloth with a mild soap solution 1 cc of liquid dish soap to one liter of clean filtered water filtered below 5 microns CAUTION DO NOT USE ALCOHOL OR ANY SOLVENTS ON ANY EXTERNAL SURFACE OF THE...

Страница 33: ...from Reichert use only the part number listed above Note Any fuse meeting the specifications listed in the Specifi cations section of this manual may be used in the device 4 Push the Fuse Holder into...

Страница 34: ...pen so that the wide Tabs are aligned with the opening at the top of the Pen Holder Refer to Figure MM 08 4 The Marking Pen can be removed by pulling it through the bottom of the Pen Holder Refer to...

Страница 35: ...ERROR SOLUTION Paper Empty Replace the Printer Paper Printer Cover Open Close the Printer Cover Printer Overheated Printer is overheating Allow printer time to cool off EEPROM Failure Abnormality of...

Страница 36: ...er Measuring Wavelength 525 nm UV Transmittance Measuring Wavelength 375 nm Measurable Lens Unprocessed Lens diameter 100mm and Framed Lens Single Lens multifocal lens progressive lens Single Focus Ha...

Страница 37: ...esidues At the end of its product life follow your local laws and ordinances regarding the proper disposal of this equipment Software Revision The software revision can be obtained by contacting Reich...

Страница 38: ...est Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class B The LensChek Plus Pro uses RF energy only for its internal function Therefore its RF emissions are very low an...

Страница 39: ...00 4 2 Floors should be wood concrete or ceramic tile If floors are synthetic the r h should be at least 30 8kV Contact 2kV 4kV 8kV 15kV Air 8kV Contact 2kV 4kV 8kV 15kV Air Compliance Level Surge IEC...

Страница 40: ...ment marked with the following symbol Radiated RF IEC 61000 4 3 80 MHz to 2 5 GHz 3V m 80 MHz to 2 7 GHz 10V m E1 3 V m E1 10 V m Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2...

Страница 41: ...maximum output power of the communications equipment Max Output Power of Trans mitter W Separation m 150kHz to 80 MHz d 3 5 V1 Sqrt P Separation m 80 to 800 MHz d 3 5 E1 Sqrt P Separation m 800MHz to...

Страница 42: ...Use a shield wire for a connecting cable to protect the output data from noise When the measurement is output using RS232C use a straight cable D sub 9 pins male D sub 9 pins female for connection The...

Страница 43: ...ng of the product Patent Warranty If notified promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Reichert will defend such acti...

Страница 44: ...Email reichert information ametek com www reichert com Authorized European Representative AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Email inf...

Отзывы: