That is, the activation sequence of
channel Ch1 and Ch2. The possible
selections will be simultaneous or
alternated. Once defined the
TREATMENT time, by pressing
again the FUNCTION key within 6
seconds, the symbol will blink on the
display. Through the
T
up
/ T
down
keys you can then change the
synchronism between
SIMULTANEOUS and
ALTERNATED
.
If you will not change
the value within 6 seconds, it will
re-assume the original function
(pulse width).
Modify the ACTION time
The ACTION time of whichever
program is preset or defined by the
USTIM system, it is not required to
modify it; nevertheless, you can
change it (temporarily in the preset
37
- Increase the stimulation
intensity,
until a good but
comfortable muscular contraction
has been achieved.
- After 2-3 minutes, gradually
increase the frequency value,
until the target value.
The target
frequency value must be
previously defined, according to
the therapeutical aim or to the
session finality.
The basic concepts about the
frequency selection have been
explained in the page 8. It is also
possible to take a suggestion from
the table of preset programs in the
page 66.
If you want simply toning
a muscle, the maximum
frequency will be 30-35 Hz. If the
aim is the hardening, you can
attain 40-50 Hz. Willing to
develop force, you can attain
60-70 Hz.
USTIM - User’s manual Rev. 2(10/19/2012)
34