Resuming the practical
operations
Whenever the unit is switched-on, the
last used program will be loaded.
Willing to apply the same program,
simply place the electrodes and
setting the intensity level of both
channels.
Select another program
Press the
PROG
button for selecting
one of the programs in memory (P1
to P17 or M0-M1-M2-M3).
To switch
between the MANUAL programs (M)
and the PRESET programs (P),
press and hold down the PROG
button for at least 3 seconds.
Modify the frequency
Press the
Hz+/Hz-
buttons
to change
the frequency (pulse rate) of the
selected program.
Modify the pulse width
Press the
T up/T down
buttons to
modify the pulse width of the selected
program.
35
Either the Frequency or the Pulse
width can be whenever changed and
the changed effect will be immediately
felt by the user.
Through the
FUNCTION
button, it is
possible sequentially selecting some
other parameters; they can be then
modified by the buttons
T up/T down
let’s see them in detail:
Modify the TREATMENT time
Press once the
FUNCTION
key, to
select the TREATMENT time. You
can change it within 6 seconds, by
using the
T up/T down
keys. After 6
seconds, the T up/T down keys will
re-assume the original function (pulse
width).
In the MANUAL programs the change
will be retained, while any change in
the PRESET programs will be lost.
Modify the SYNChronism between
Ch1 and Ch2
(program M0 only)
Only if you are using the program M0,
you will be allowed to modify also the
synchronism between the channels.
36