Refine PT-X Скачать руководство пользователя страница 9

9

the temperature is-20℃ - +40℃ .

6.2.4 When the device is not in use, turn off the power supply and unplug the power plug. If it is not used for a long time, it should 
be energized and connect to water and air once a month for five minutes.

6.3 maintenance

6.3.1  Air filter

1) When water accumulate in the filter, turn the knob at the bottom of the filter counterclockwise to drain the water, and tighten the 
knob clockwise.
2) Replacement of the filter element: Use a filter wrench to unscrew the transparent cover of the air filter, then use the wrench to 
unscrew the black nut at the lower end of the filter element, remove the white filter element and discard it into the trash can, re- 
place it with a new filter element, and reinstall the black nut and transparent shell. It is recommended to replace the filter element 
every 24 months, and the spare filter element is included in the accessory.

Figure 10 Schematic diagram of filter element replacement

7 Environmental Protection 

The instrument contains no hazardous components and can be disposed of and destroyed in accordance with the relevant local 
regulations.

Part

Toxic or harmful substances or elements

Pb

Hg

Cd

Cr6+

PBB

PBDE

Main unit

andpiece

Tip

Nozzle Foot pedal

Mechanical elements, including bolts, 
nuts, washers, etc.

○:Indicates that the content of the toxic substance in all homogeneous materials of the part is below the limit requirement 

stipulated in SJ/T-11363-2006 Limit Requirements for Toxic and Hazardous Substances in Electronic Information Products.
X:indicates that the content of the toxic substance in at least one of the homogeneous
materials of the part exceed the limit requirement specified in SJ/T-11363-2006. Please dispose according to the local laws or 
consult with dealer from whom you purchased it about waste disposal.

8 After service 

8.1 After the equipment is sold, the manufacturer will be responsible for quality problem according to the warranty card. For specific 
items, please refer to the warranty instructions in the warranty card.
8.2 This product does not contain self-maintaining parts. All maintenance, adjustment, calibration, and modification of technical 
parameters of the product can only be carried out by the technicians or special repair shops. If the customer needs to repair by 
himself, the manufacturer can provide circuit diagrams, component lists, legends, calibration rules, or other information necessary to 
help the user's qualified technicians repair the equipment parts designated by the manufacturer, but the manufacturer will not bear 
the consequences arising therefrom.
8.3 The user must use the original accessories, please contact your local dealer or the manufacturer to purchase. It is forbidden to use 
accessories of other brands to avoid damage to the equipment or other dangers.
8.4 After the handpiece, tips and other accessories are damaged, users should not repair them by themselves. Please purchase new 
parts and replace them before use. If you need relevant information, please contact the manufacturer.

9 EMC-Declaration of conformity

Notice:1) Without the express consent of Guilin Refine Medical Instrument Co., Ltd., unauthorized changes or modifications to the 

equipment may cause electromagnetic compatibility problems of this equipment or other equipment.
2) The design and test of the PT X dental scaler conform to the operating regulations of electromagnetic compatibility. Its basic 
performance is: the tip installed at the front end will vibrate when working, water will spray out, and the front end of the handpiece 
will emit light at the same time .
3) Note: During the electrical fast transient burst test, there may be situations in which the power cannot be adjusted due to the 
interference of the touch buttons. This situation will not affect the power output of the whole machine, and it can be restored by 
itself after the test is completed. Consultation A clinician with professional training experience and field including the use of special 
equipment or systems to determine that the risk is acceptable.

Warning:

1) The equipment or system should not be used close to or stacked with other equipment. If it must be used close or stacked, it should 
be observed to verify that it can operate normally under the configuration used.
2) The operation of the PT X dental scaler at a value lower than the minimum amplitude or minimum value specified in this manual may 
lead to inaccurate consequences.
3) Except for transducers and cables sold by equipment or system manufacturers as spare parts of internal components, the use of 
accessories, transducers and cables outside the regulations may result in increased emission of the equipment or system or reduced 
immunity.
4) The use of accessories, transducers or cables outside of the regulations together with equipment and systems may result in an 
increase in the emission of the equipment or system or a decrease in immunity.

Guidance and manufacturer’s declaration of electromagnetic emissions

The model PT X is intended for use in the electromagnetic environment specified
below. The customer or the user of the model PT X should assure that it is used in such an environment.
Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1

The model PT X uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely to cause 
any interference in nearby electronic equipment.

Содержание PT-X

Страница 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Страница 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Страница 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Страница 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Страница 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Страница 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Страница 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Страница 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Страница 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Страница 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Страница 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Отзывы: