2

20. This equipment has electromagnetic interference. Do not use it around electronic surgery, and be cautious in an environment with 
strong electromagnetic interference when using this equipment.
21. This equipment does not contain toxic or hazardous substances, and it can be discarded in accordance with the relevant laws and 
regulations on discarded medical devices.
22. Do not step on the foot pedal when the cable of the air polish handpiece is removed from the main unit.
23. Before using the ultrasonic handpiece, please make sure that the cable of the air polish handpiece is correctly placed on the handle 
bracket; similarly, before using the air polish handpiece, please make sure that the cable of the ultrasonic handpiece is correctly placed 
on the ultrasonic handle bracket.
24. The tip must be tightened.
25. The air polishing handpiece nozzle should not be aimed at people.
26. There might be injury if the powder accidentally sprays into the eyes. We strongly recommend that all personnel (doctors, nurses, 
patients) wear goggles during air polishing treatment.
27. Please do not unload the powder tank when stepping on the pedal or when the machine is working. 
28. Before replacing the air polishing handpiece or the nozzle, please use a syringe to blow the moisture at the joints at both ends 
(especially the gas interface) to prevent moisture from entering the gas path and avoid clogging of the powder in the pipeline.

Contraindications

1. The hemophilia patient is forbidden to use this equipment     2 . The patients or the doctors with heart pacemaker are forbidden to 
use this equipment.    3. Heart disease patients, pregnant women and children should be cautious to use the equipment.   4. Patients 
with respiratory diseases such as asthma and chronic bronchitis are not allowed to use this device.

Symbol Instruction

Symbol

Implication

Symbol

Implication

Symbol

Implication

Trademark

See instruction mannual

Grounding protection

Warning and important 
notes! See the attached 
documents

Anti-drip equipment

Sterilizable spare parts under 
high temperature 

Recycling

Manufacturer

Indoor only 

Date of manufacture

BF type applied part

Alternating current

Power-on button

Power-off button

ON/OFF

Power switch

IPXO

Ordinary equipment

Power adjustment

Foot pedal plug

Storage condition, 

temperature limit: -20℃ 
- +40℃

Storage condition, air pressure 
limit: 70kPa - 106 kPa

Storage condition, humidity 
limit: 10% - 93%

Keep dry

Fragile

Compliance with WEEE 
directive, Discarded as per 
regulations

Boost mode

Anhydrous mode

Clean mode

Forward

Guilin Refine Medical Instrument Co., Ltd. is a high-tech enterprise integrating R&D, production and sales of dental products, with a 
complete quality assurance system.

1 Product Introduction 

1.1 Brief introduction

PT X dental scaler combines ultrasound scaler and air polishing scaler. The main features are: 
1) According to the selected handpiece, automatically switch the working mode.
2) Touch LCD screen.
3) Special chemical solutions such as hydrogen peroxide, sodium hypochlorite and chlorhexidine can be used to improve the clinical 
treatment effect.
4) LED handpiece enables convenient Clinical operation.
5) The automatic frequency tracking system for the best working condition, which brings more stable performance of device.
6) Detachable ultrasonic handpiece and air polishing handpiece can be sterilized under high temperature of 134 and high pressure 
of 0.22MPa.

1.2 Model: PT X
1.3 Scope of application

This product includes an ultrasonic system and an air polish system. The ultrasonic system is used for periodontal treatment in oral 
clinical treatment to remove tartar and plaque, root canals cleaning; Air polish system is used to remove plaque and pigment, it can 
also used for implants maintenance.

1.4 Device configuration

 Please refer to the packing list for device configurations.

1.5 Structure and components

The machine consists of main unit, handpiece (ultrasonic handpiece, air polishing handpiece), cable, bottle, tips, dental air polishing 
powder, torque wrench and foot pedal, Power cord.

1.6 Replacement instruction for main accessories.

Name

Model

Replacement instruction

Air polishing andpiece

XP-1 / XP-2

Reusable and Durable; Please change a new handpiece when the handpiece is 
blocked and cannot be dredged, or the sealing performance is reduced, or the 
ventilation pipe is seriously worn out

Содержание PT-X

Страница 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Страница 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Страница 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Страница 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Страница 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Страница 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Страница 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Страница 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Страница 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Страница 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Страница 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Отзывы: