Refine PT-X Скачать руководство пользователя страница 5

5

C, it will automatically return to the previously set grade.

Figure 9 
Schematic diagram of multi-function foot pedal

3.2 Ultrasonic system

3.2.1 Scaling
1) Turn on the power switch and pick up the ultrasonic 
handpiece, afterwards the panel will automatically shift 
to Ultrasonic mode.
2) Press the G button to enter Scaling mode.
3) Select the tip and tighten it on the handpiece by 
torque wrench.
4) When stepping on the button A, tip vibrates with LED 

on and water emits; Release the button, vibration and water stop, LED will be off 10 seconds later. 
5) The handpiece is generally held in a pen-holding position.
6) The frequency is extremely high when the machine is on working. It can remove the tartar with the tips softly stroke on the teeth 
surface like erasing motion. DO NOT stay long or put overmuch pressure on the teeth. 
7) Vibrating intensity: Adjust the vibrating intensity according to your need. The setting power is level 3, please adjust the vibrating 
intensity according to the teeth sensitivity and hardness of calculus during clinical application.
8) Water volume: Please adjust the volume on the panel.
9) During clinical scaling, please keep the side of the tip adapt to the tooth surface horizontally, with light lateral pressure to allow the 
tip vibrate freely.
10) After operation, please keep device working for 30s with water to flush the handpiece and tips; Remove the tip and get it sterilized. 
3.2.2 Periodontal treatment
1) Use a torque wrench to tighten the periodontal treatment tip to the ultrasonic handpiece. Click the P button on panel to enter the 
periodontal treatment mode.
2) The operation and adjustment methods are similar to Scaling mode.
3.2.3 Endodontic irrigation
1) Tighten the endo file to the handpiece by endo wrench.
2) Click on the E button and enter the Endo mode. 
3) The default power is level 1 under Endo mode, operator can adjust the power according to the need in clinical treatment.
4) Select the appropriate Endo file and slowly place it into the root canal of the patients teeth. Step on the foot pedal to start 
ultrasonic endodontic irrigation.
5) During clinical cleansing, the file should not be pressed too tightly when it is in the root canal 
6) Do not step on the pedal before the file is in the root canal. 
7) The suggested power range for endodontic irrigation is level 1-level 5.

3.3 Air polishing system

1) Add appropriate amount of powder to the powder tank (the amount of powder added should be between "Max" and "Min" marked 
on the the tank), then tighten the upper cover of the powder tank, and fix the powder tank on the main unit.
2) Pick up the Air polishing handpiece, afterwards the panel will automatically shift to Air polishing mode.
3) Adjust the water volume and air pressure, aim the nozzle at a pool, press the foot button A, and confirm that the nozzle can spray 
gas, powder and water spray normally before use.
4) Before Air polishing treatment, please wear goggles and veil on the face of the patient. Users should wear goggles or a protective 
mask.
5) The handpiece is generally held in a pen-holding position.
6) Adjust the water volume and air pressure to an appropriate level. The recommended water volume starts at level 5 and the air 
pressure starts at level 3. According to the sensitivity of the patient's teeth and hardness of the dental plaque, adjust the water 
volume and air pressure at any time during the clinical process; increasing the air pressure will enhance cleaning effect, but will 
weaken the polishing effect; increasing the amount of water will enhance the polishing effect, but will weaken the cleaning effect.
7) The nozzle should be aligned with the tooth surface during the cleansing, but avoid direct contact. Keep the nozzle and the tooth 
surface at a distance of 3-5mm at 30-60 angle. The smaller the angle, the larger the cleaning area; during the cleansing, please 
Perform a small circular motion on the surface of the tooth.
8) The air/powder mixture reflected from the tooth surface should be evacuated by using a strong suction device on the dental unit 
during treatment.
9) After treatment, adjust the water volume to the maximum level, and polish the surface of teeth.

3. 4 Cleaning Mode

It is recommended to flush and disinfect the pipeline of the device every day. "Cleaning" mode allows cleaning and disinfection of the 
pipeline to reduce the accumulation of crystals and the number of bacteria in the pipeline.
1) Fill a water bottle with distilled water or demineralized water.
2) Pick up the ultrasonic handpiece, point the handpiece at the sink, click the Cleaning button on the screen, and press the button D 
on foot pedal to start cleaning the pipeline. Then, the pedal can be released.
3) After cleaning for 30 seconds, the device will automatically stop the Cleaning mode. You can also press the button D on foot pedal 
again in the Cleaning mode or click Purge on the screen to stop cleaning.
4) After cleaning, put the ultrasonic handpiece back into the bracket. And then, pick up the air polishing handpiece, point the 
handpiece nozzle at the pool, and click the Cleaning button again, so that the device will automatically blow out the residual powder 
in the pipeline and release the high pressure gas in powder tank.
5) After cleaning for 20 seconds, the device will automatically exit the Cleaning mode. You can also click Purge on the screen to stop 
cleaning.

3.5 Setting

1) Click "ON" /"OFF" to turn on or off the heating function. (According to the actual water output, the water temperature is different).
2) Click the "Settings" button at the bottom right corner of the screen to enter the "System Settings" interface
3) Click Calibration button, the system will automatically calibrate the air pressure to the value set in the program
4) The Bluetooth pedal has been matched to the machine, any subsequent replacement of foot pedal or main unit needs to make 
new pair.  

Содержание PT-X

Страница 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Страница 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Страница 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Страница 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Страница 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Страница 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Страница 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Страница 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Страница 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Страница 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Страница 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...