10

RF emissions CISPR 11

Class B

PT X is suitable for being used in domestic establishment and in 
establishment directly connected to a low voltage power supply 
network which supplies buildings used for domestic purposes 

Harmonic emissions IEC 61000-3-2

Complies

Voltage fluctuations/ flicker emissions IEC 
61000-3-3

Complies

Guidance & Declaration - electromagnetic immunity

The model PT X is intended for use 1n the electromagnetic environment specified
below. The customer or the user of the model PT X should assure that It is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment 
guidance

Electrostatic 
discharge (ESD) IEC
61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete 
or ceramic tile If floors are covered 
with synthetic material, the relative 
humidity should be at least 30 %.

Electrical fast 
transient burst IEC 
61000-4-4

± 2 kV for power supply lines
± 1 kV for Input/Output lines

± 2 kV for power supply lines,  
± 1 kV for interconnecting cable 

Mains power quality should be that 
of a typical commercial or hospital 
environment.

Surge
IEC 61000-4-5

± 1 kV line to line
± 2 kV line to earth

± 1 kV line to line

Mains power quality should be that 
of a typical commercial or hospital 
environment.

Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines IEC 61000-4-11

<5% UT (>95% dip in UT) for 0.5 

cycle
40% Ur (60% dip in UT) for 5 cycles
70% UT (30% dip in UT ) for 25 
cycles

<5% UT (>95 % dip 1n UT ) for 5 

sec

<5% UT (>95% dip in UT .)

for 0.5 cycle
40% UT (60% dip in UT for 5 
cycles
70% UT (30% dip in UT for 25 
cycles

<5% UT (>95 % dip in UT)

for 5 sec)

Mains power quality should be that 
of a typical commercial or hospital 
environment If the user of the model 
PT X require continued operation 
during power mains interruptions, it 
is recommended that the model PT X 
be powered from an uninterruptible 
power adapter or a battery.

Power frequency 
(50/60 Hz)
magnetic field
IEC 61000-4-8

3A/m

3A/m

Power frequency magnetic fields 
should be at levels charactenstic 
of a typical location in a typical 
commercial or hospital environment

NOTE Ur is the ac mains voltage prior to application of the test level.

Guidance & Declaration - Electromagnetic immunity

The model PT X is intended for use in the electromagnetic environment specified below. The customer or the user of the model PT X 
should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Conducted RF IEC 
61000-4-6
Radiated RF IEC 
61000-4-3

3Vrms
150kHz - 80MHz
3V/m
80MHz - 2.5GHz

3Vrms
3V/m

Portable and mobile RF communications equipment should 
be used no closer to any part of the model PT X, including 
cables, than the recommended separation distance 
calculated from the equation applicable to the frequency of 
the transmitter.
Recommended separation distance
d=[3,5/V1]×P

1/2

d=1.2×P

1/2

 80 MHz - 800 MHz 

d=2.3×P

1/2

 800 MHz - 2.5 GHz

Where P is the maximum output power rating of the 
transmitter In watts (W) according to the transmitter 
manufacturer and d Is the recommended separation distance 
in meters (m).
Field strengths from fixed RF transmitters, as determined by 
an electromagnetic site survey, ª  should be less than the 
compliance level in each frequency range. 

b

interference may occur In the vicinity of equipment marked 
with the following symbol:

NOTE 1 At 80 MHz - 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people.
A) Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the model PT X is used exceeds the applicable RF compliance level above, the model PT X 
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such 
as reorienting or relocating the model PT X.
B) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Recommended separation distances between

portable and mobile RF communications equipment and the model PT X

The model PT X is intended for use in electromagnetic environment in which radiated RF disturbances is controlled. The customer or 
the user of the model PT X can help prevent electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the model PT X is recommended below, according to the maximum output 
power of the
communications equipment.

Rated maximum output power 
of transmitter W

Separation distance according to frequency of transmitter m

150kHz ~ 80MHz      d =1.2 

80MHz ~ 800MHz   d =1.2 

800MHz ~ 2.5GHz      d =2.3 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

Содержание PT-X

Страница 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Страница 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Страница 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Страница 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Страница 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Страница 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Страница 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Страница 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Страница 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Страница 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Страница 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Отзывы: