Dental Scaler Instruction Manual

Precautions

Warning: If you neglect these precautions, it might cause personal injury such as electric shock, fire or damage to the product.   

1. This product is not a home appliance. It is only applicable for hospitals and dental clinics. The use of the product must comply with 
the requirements of the relevant operating specifications and relevant laws and regulations of the medical department. The user must 
be a professionally trained and qualified dentist or technician. Adequate protection (such as goggles, mask, etc.) should be equipped 
during operation to prevent cross-infection. 
2.Please use an independent power outlet with protective grounding. Do not unplug the power cord with wet hands.
3. Please fully insert the power plug of the device into the power socket, please do not use other power source other than the 
specified voltage.
4. Do not damage, modify, pull, bend or twist the power cord excessively, and do not place heavy objects on the power cord.
5. After using the machine, turn off the power switch and pull out the power plug.
6. When the equipment recovers after power off, please await the equipment to be stable before stepping on the pedal. The operator 
should release the pedal in time when the power is off.
7. Please do not place the product on an unstable workbench, such as a swaying table, an inclined surface, or a location subject to 
vibration.
8. This equipment is a reusable product. Keep the equipment clean before and after use. Before each use, the handpieces, tips, torque 
wrench etc. must be disinfected and sterilized. It is suggested to disinfect and sterilize according to the recommended method in 
Chapter 5 of this manual.
9. Do not knock or scrape the handpiece. Do not pull the cable forcefully during the use of the equipment to avoid damages.
10. Do not bend or polish the tips. The tips must be screwed to the handpiece with a torque wrench, and water spray must be 
generated during operation.
11. Before each operation, you should work outside the patient's mouth for more than 10 seconds under the condition of water to 
drain the water in the pipeline at the back of the handpiece
12. When the tip is damaged or worn out, the vibration intensity might decrease. The operator should replace a new tip in time 
according to the clinical situation. It is not recommended to use the tip after the abrasion of tip exceed 2mm or more. You can use the 
tip comparison indicator to check. If the length is out of the first line, the power remains the same; if it is between the first and second 
line, the power would decreased to 80%; if it reaches or surpass the second line, the power would drop massively, on the occasion, 
replacement of tip is recommended.
13. Do not screw or unscrew the tip while stepping on the foot switch, or when the machine is on working.
14. The inner thread of the tip manufactured by certain manufacturers are rough, rusty and might break the thread system, which will 
cause irreparable damage to the scaler. Please use corresponding tips of Refine brand.
15. Choose the appropriate power according to different types of tips. See Annex Tip power chart There are  recommended power and 
water volume for different types of tip in the system. You can double -click the current mode to choose the tip.
16. Do not use unclean water
17. Improper cleaning and treatment of titanium implants, porcelain restorations, etc. can easily cause loosening of the adhesive, 
cracking of the porcelain restorations, or even cracking of the porcelain. The cleaning or treatment of the oral cavity of such patients 
should be carefully considered.
18. Noise will be generated during the scaling process. Those who are sensitive to noise can wear earplugs by themselves.
19. We will be responsible for the safety only when the maintenance, repair and modification of the machine are carried out by the 
company or its authorized distributors, the replacement parts are from the company, and the operation is on the basis of instruction 
manual. 

Содержание PT-X

Страница 1: ...torque wrench and water spray must be generated during operation 11 Before each operation you should work outside the patient s mouth for more than 10 seconds under the condition of water to drain the...

Страница 2: ...ng Manufacturer Indoor only Date of manufacture BF type applied part Alternating current Power on button Power off button ON OFF Power switch IPXO Ordinary equipment Power adjustment Foot pedal plug S...

Страница 3: ...inst electric shock BF type 4 Degree of protection against harmful ingress of water Ordinary equipment IPX0 Foot pedal is anti drip device IPX1 5 Degree of safety application in the presence of a flam...

Страница 4: ...cover and install 2 AA batteries Switch on the main unit and the foot pedal can be connected and controlled when leaving the factory the Bluetooth foot pedal and the main unit have been paired and con...

Страница 5: ...s powder and water spray normally before use 4 Before Air polishing treatment please wear goggles and veil on the face of the patient Users should wear goggles or a protective mask 5 The handpiece is...

Страница 6: ...l Poor connection under wired pedal mode Connect the wire and screw it well Out of battery under wireless pedal mode Replace a new battery The nozzle does not spray gas but there is water spray after...

Страница 7: ...clean running water desalted water distilled water or deionized water for at least 5 times with duration of no less than 60 seconds for each time Check whether the cleaned parts are clean or damaged I...

Страница 8: ...of the Ultrasonic handpiece reaches the specified service life number of times please replace it in time 5 Do not use the machine when the machine is being cleaned disinfected sterilized 5 6 Packaging...

Страница 9: ...l accessories please contact your local dealer or the manufacturer to purchase It is forbidden to use accessories of other brands to avoid damage to the equipment or other dangers 8 4 After the handpi...

Страница 10: ...test level Compliance level Electromagnetic environment guidance Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150kHz 80MHz 3V m 80MHz 2 5GHz 3Vrms 3V m Portable and mobile RF communicati...

Страница 11: ...The final interpretation right belongs to Guilin Refine Medical Instrument Co Ltd Refer to the product packaging label for the product production date After service and warranty instruction 1 Period v...

Отзывы: