5. GENERAL SAFETY  

Before turning on the device for the first time, read this manual carefully. 

ƒ

 

The device may only be used under the supervision of the physician prescribing the treatment. 

ƒ

 

Do not use on bare skin. It is recommended to utilize a protective sheath for treatment

ƒ

 

All personnel must be properly trained before using the device. This training should include the 
servicing technique, maintenance, and verification of proper functioning and safety precautions 
associated with using the device. 

ƒ

 

Shockwave therapy requires bodily contact, therefore do not use it on damaged skin 

ƒ

 

Do not apply shockwave therapy: To areas of the body with gas-containing organs, close to large 
nerve bundles or blood vessels, on the spinal cord, or on the head. 

ƒ

 

We do not recommend applying shockwave to the body parts under local anesthesia. 

ƒ

 

The device does not use any drugs, creams, gels, or other substances. 

ƒ

 

The device is equipped with a protection system that prevents the connection of accessories 
other than those supplied by the manufacturer. It will not operate with equipment from other 
manufacturers. 

ƒ

 

Only the power supply adapter approved and supplied by the manufacturer, complying with the 
specifications of the machine, may be connected to the device. 

ƒ

 

The product complies with Class A emissions according to CISPR11, IEC 60601-1-2. 

ƒ

 

Portable and mobile high-frequency communication devices (such as mobile phones) may affect 
the function of the device. 

ƒ

 

The electrical cables which are to be connected to the device must be installed and tested 
according to the existing valid standards (IEC 364). If it is not known whether cables are safe 
and/or correct, then they should be checked and/or upgraded by an inspection engineer. 

ƒ

 

Check whether the parameters of the main power supply correspond to the requirements of the 
device according to the machine’s specification. 

ƒ

 

The device requires the environmental conditions that are stated in the Technical Parameters 
section. It must not be used in an environment where there is a danger of explosion or of water 
penetrating the device. The device cannot be in connection with flammable anesthetics or 
oxidizing gasses (O2, N2O, etc.). 

ƒ

 

Do not place the device in direct sunlight or near strong electromagnetic fields to prevent 
mutual functionality influence. If this happens, move the device further away from the source of 
interference or contact an authorized service department of our company. 

ƒ

 

Inspect the device thoroughly before each use. Look for loose cables, cracked cable insulation, 
cracks in the shockwave applicator’s housing and functional behavioral differences in the display 
or the operating elements. If any anomalies or inconsistencies are found, stop using the device 
and contact an authorized Refine USA service department. If the behavior of the device shows 
any divergences from the functionality procedures described in this user’s manual, stop using 
the device and contact an authorized member of the service department. 

ƒ

 

If the device shows any defects or if there are any doubts concerning its correct and safe 
functioning, terminate therapy immediately. If the source of the concern can be determine after 
a thorough study of the user's manual, then contact an authorized service department 

Содержание APEX MAGNUM

Страница 1: ...2019 APEX MAGNUM DUAL FUNCTIONING ACOUSTIC WAVE THERAPY SYSTEM Revision 1 12 11 2018 User s Guide...

Страница 2: ...ion 9 2 3 Pneumatic Accessories 10 2 4 Electromagnetic Accessories 10 2 5 Replacement Parts 11 2 6 Pneumatic Transmitter Adapter Replacement 12 2 6 1 Pneumatic Piston Replacement 12 2 7 Electromagneti...

Страница 3: ...ON WITH THIS DEVICE Only use BABY OIL IF DESIRED The use of gels or lotions render both hand pieces inoperable The damage from the use of gels or lotions is not covered under the warranty Contact Refi...

Страница 4: ...panel USER S INSPECTION DAMAGE CAUSED IN TRANSIT The unit and all accessories must be checked for potential defects transport damage or loss immediately after receipt Specifications are subject to cha...

Страница 5: ...g feature Based on detailed research and practical use of the device the well organized predefined programs will provide recommendations for the treatment of various conditions The device consists of...

Страница 6: ...ive pressure amplitude 1 to 10 MPa Short time duration 1 mu to 20 mu Rapid pressure increase 100 ns Broad frequency spectrum 1 Hz to 1 MHz For therapeutic applications these values are lower especiall...

Страница 7: ...corticoid therapy Shockwave application can cause undesirable heart activity 1 4 INTENDED USE FOR SHOCKWAVE TREATMENT ESWT is for the intended use as a treatment for musculoskeletal pain 1 5 CONTRAIN...

Страница 8: ...ug 8 Filter cup with pressure gauge 9 Plug for pedal switch 10 Plug for Electromagnetic wave handle 2 1 ACCESSORIES LIST Device 1pc A shock wave handle 2pcs E Shock wave handle 2pcs A handle transmitt...

Страница 9: ...itter Adapters Pneumatic transmitter adapter treatment heads 11 Electromagnetic Transmitter Adapters Electromagnetic transmitter adapter treatment heads 5 Foot Pedal Foot pedal for power 1 Apex Silico...

Страница 10: ...to the bottom cooling vents Do not put any heat producing devices or any objects containing water or other liquids on the device Do not place the device close to appliances producing strong electroma...

Страница 11: ...Transmitter Adapters B 15 mm B 20 mm B 25 mm Convexed Pneumatic Transmitter Adapters C 15 mm C 20 mm C 25 mm Concaved Pneumatic Transmitter Adapters 2 4 Electromagnetic Accessories Electromagnetic Sho...

Страница 12: ...escription Pneumatic Handle Pneumatic hand piece Electromagnetic Handle Shockwave hand piece Electromagnetic Transmitter Adapters Shockwave handle tips Foot Pedal Foot pedal for power Pneumatic Transm...

Страница 13: ...to choose the proper transmitter Secure the transmitter on the applicator prior to treatment 1 Transmitter Adapter Screw Cap After Installation 2 6 1 Pneumatic Transmitter Piston Replacement Assembled...

Страница 14: ...ols to choose the proper transmitter Secure the transmitter on the applicator prior to treatment Unscrew adapter by twisting counter clockwise until removed from handpiece Screw chosen adapter head in...

Страница 15: ...ng to select language set time Notice ADVANCED SETTING is only for Engineering debugging DO NOT USE Click A shock to enter Interface as FIG 3 1 Click on the body part chosen for treatment on the scree...

Страница 16: ...f rounds completed 600 impacts are one round 10 Frequency 6 HZ stands for the system default work frequency is 6 HZ Energy 2 0 BAR stands for the system default work Energy is 2 0 BAR click on and to...

Страница 17: ...the default intensity frequency and number of impacts The automatic mode can also be adjusted freely to the most suitable parameters Treatment example for Elbow adipose tissue In interface as Fig 4 1...

Страница 18: ...15HZ the energy is 50 130MJ When the frequency is 16HZ the energy is 50 100MJ 3 3 Touch Screen Calibration 1 Before powering on the device press on the screen with your finger do not remove your finge...

Страница 19: ...nd piece control board o Please check if the plug of the cooling handle 8 is connected accurately Reboot device E1 Connection failure between display control board and function control board This may...

Страница 20: ...2 detergent solution to clean the exterior of the Apex Therapy System and its parts Never use cleaning agents containing alcohol ammonia benzene thinners etc Never use abrasive cleaning materials whi...

Страница 21: ...lled and tested according to the existing valid standards IEC 364 If it is not known whether cables are safe and or correct then they should be checked and or upgraded by an inspection engineer Check...

Страница 22: ...ransport under the stated conditions Before the start of therapy make sure that all set parameters comply with your requirements To terminate operation do not use the main power switch Instead press t...

Страница 23: ...ents are not covered Failure to do so may result in permanent damage Correct usage of protective sheath Correct usage of protective sheath Improper use of protective sheath Sheath covering air vents C...

Страница 24: ...gether with a photocopy of your bill of sale What Refine USA LLC will do Refine USA LLC will evaluate your report of a possible defect to determine whether a defect exists and if it does exist Refine...

Отзывы: