Instruction Manual DIGIMON-SE / DIGIMON4
English
19
9. Press the
button to calibrate the high pressure side (HP)
Repeat
from Point 6
10. Once both sensors are calibrated, press
to exit the sensor calibration
mode.
Calibrating the temperature sensors / clamps (0°C / 32°F ice water)
1. Switch
off
the DIGIMON
2. Plug in the thermocouples (K-type).
The DIGIMON and the thermocouples must remain at a constant room
temperature for at least 20 minutes. The thermocouple cables must be straight
(unwound).
3. While holding down the
button continuously, switch on the device by
pressing the
button once. Continue holding down the Light button until
the display shows, "Password"
000
4. Press the
/
buttons to scroll to code number
020
.
5. Press the
button once.
Display shows
“T-call0”
6. Press the
button once and then press the
button once to enter
temperature sensor/clamp calibration mode.
During calibration, ensure that both thermocouples are well connected to the
DIGIMON.
The display shows
(low pressure side)
SET
R
-
SET
R
-
R
+
ON/OFF
LIGHT
ON/OFF
R
-
LP
011xx
24.4°C
Temperature code (will change)
Your temperature reading (will not change)