Instruction Manual DIGIMON-SE / DIGIMON4
English
17
Calibrating pressure sensors
A correct calibration requires a certified monitor.
(e.g. REF-CLASS-GAUGE P/N 4682293)
1. Switch off the DIGIMON
2. Press down the
button and switch on the device using the
button.
The display shows "
Password
"
000
3. Press the
/
button to select Code
009
.
4. Press the
button to exit.
The display shows
"P-call"
5. Press the
button and then the
button to calibrate.
400 psi
should be displayed on the LP side display
For a correct calibration, there should be no pressure or vacuum in the
DIGIMON.
6. Press the
button to select the low pressure side's calibration pressure
(400psi).
7. Press the
button to exit.
8. Wait 3-5 seconds until the pressure is stable and press the
button
again.
The display shows
"0000"
9. Pressurise the DIGIMON with a calibration pressure of 400psi (27.58 bar).
When the pressure is stable, press the
button to confirm.
The display shows
"LP End"
10. The calibration of the low pressure side has been completed.
11. Press the
button to calibrate the high pressure side (HP) Repeat from
Point 6 or press the
button to end the calibration.
ON/OFF
R
-
SET
SET
SET
R
+
SET
R
-
SET
R
-
R
+
ON/OFF
LIGHT