8
Operating Instructions
Power ON/OFF
Press the
POWER
button to turn the instrument on. To turn the instrument
off, press and hold the
POWER
button.
Automatic Power-Off
As a default the instrument will automatically turn off after 15 minutes of
inactivity. Follow the steps below to Enable/Disable the Auto-Power
Off function.
1. Turn the meter off.
2. Press the
ON/OFF
button (to display the full screen).
3. Quickly press the
mA%
button when the instrument is in the mainte-
nance state as indicated by "1.8.8.888".
4. "AP-XX" will appear on the display.
5. Press the button to toggle between "AP-ON" & "AP-OF". "AP-OF"
indicates that the automatic power-off function is disabled, while
"AP-ON" indicates the automatic power-off function is enabled.
6. Press the
mA 100%
button to save the required setting.
7. Press and hold the
ON/OFF
button to exit the maintenance state and
turn the instrument off.
Output Function
The output terminal of the instrument can produce DC voltages set by the
user or simulating resistance.
Do not apply any voltage to the output terminal during the operation. If any
improper voltage is applied to the output terminal, it will cause damage to
the internal circuits.
continued...