10
25% Step Current Output
1. Connect the test leads to the meter as stated in the Current
Output procedures.
2. Press the
V/mA
button to select mA function and mA will appear on
the display.
3. Press the
mA 25%
button and will appear on the display.
4. Press the
mA/%
button to select the set output mA or % and mA or
mA% will appear on the display.
5. Press the
button to change the output value in 25% increments,
in which 0% indicates 4mA and 100% indicates 20mA.
6. Press the
ON/OFF
button to turn the output signal on and off as
indicated by ON or OFF on the LCD display.
7. Press the
mA 25%
button to exit the 25% current output setup.
Current Output Set for Zero Point & Full Scale
1. Connect the test leads to the meter as stated in the Current
Output procedures.
2. Press the
V/mA
button to select the mA function and mA will appear
on the display
3. Press the
mA 100%
button. , 0 and FS will appear on the display.
4. Press the
mA/%
button to select the set output mA or % and mA or
mA% will appear on the LCD display.
5. Press the
buttons to change the output values at 100%, in
which 0% indicates 4mA and 100% indicates 20mA.
6.
Press the
ON/OFF
button turn the output signal on and off as indicated
by ON or OFF on the display.
7. Press the
mA 100%
button to exit the full scale current output set up.
continued...