15
Data Logging
Data Recording Modes
Manual Data Logging:
Press the
LOG
button to manually log up to
99 readings on a SD memory card (see
Manual Data Logging Mod
e for
full setup instructions).
Automatic Data Logging:
Setup the meter parameters in order to
automatically log data on a SD memory card. The number of data points
is limited by the size of the memory card.
Note:
it is recommended to use a class 4 SDHC memory card between 1Gb
and 16Gb. insert the SD memory card in the slot at the bottom of the meter.
The card must be inserted with the label side facing the rear of the meter.
Manual Data Logging Mode
In manual mode, the
LOG
button is pressed to manually log a reading on
the inserted SD card at any time. In order to setup the meter for manual
data logging, follow steps 1 through 7 below.
1. Set the data logging sampling rate to "0" seconds (see
Setting the
Data Logging Sampling Rate
section for details).
2. Press the
REC
button and "REC" will appear on the LCD.
3. Press the
LOG
button to enter manual data logging function as
indicated by "DATALOGGER".
4. The meter will also display P-X (X = memory position number
between 1 and 99).
5. Press the and buttons to select one of the 99 data memory
positions in which to record.
6. Press the
LOG
button to save a reading to memory. 'DATALOGGER'
will flash each time a data point is stored.
Note:
If a card is not inserted or the card is defective, the meter will flash
"EMPTY". in this case, power the meter OFF and try again with another
SD memory card or verify that the card is correctly inserted.
7. To exit manual data logging mode, press and hold the
REC
button
2 seconds to resume normal operation.
continued...