Selecting the Temperature Unit of Measure (t-CF)
Follow steps 1 through 3 when "t-CF" appears on the LCD.
1.
Press the and buttons to select to select between °C and °F.
2. Press the
ENTER
button to confirm selection.
3. Press the
SET
button to save and skip to next parameter.
Note
: At any time, you can press the
ESC
button to exit the Setup mode
and resume normal operation.
Setting CD Temperature Compensation Factor (PEr C)
This function is only for the Conductivity (TDS) mode and is intended to
adjust the probe's temperature compensation value in % per °C. Follow
steps 1 through 3 when "PEr C" appears on the LCD.
1.
Press the and buttons to select the desired temperature
compensation. The default value is 2% per °C.
2. Press the
ENTER
button to confirm selection.
3. Press the
SET
button to save and skip to next parameter.
Note
: At any time, you can press the
ESC
button to exit the Setup mode
and resume normal operation.
ESC
Press the
ESC
button when "ESC" appears on the LCD to exit the Setup
mode and resume normal operation.
Verify Set Time and Date
During normal operation, press the
TIME
button once to display the hour,
minute, seconds. Press the button again to display the year, month and day.
Verify Sampling Rate
During normal operation, press the SAMPLinG CHECk button to view the
selected sampling rate.