12
0°C Calibration
Insert temperature probe into a 0°C standard calibrator unit
before calibration. Hold the probe until condition is stabilized
before starting calibration.
While the meter is off,
press and hold the
POWER
and BACKLIGHT
buttons for 2 seconds to enter the setup function. Press the
REL
button
to enter the calibration function. Enter the Pt selection first. Press the
°C/°F
button to confirm the Pt selection. Press the
REL
button to enter
the calibration function. Press the
REL
button to confirm the current
calibration value. Press the
°C/°F
button to confirm the selection and
resume normal operation.
Reset Calibration Value to Factory Settings
While the meter is off,
p
ress and hold the
POWER
and BACKLIGHT
buttons for 2 seconds to enter setup function. Press the
REL
button
to enter calibration mode. Enter Pt selection first by pressing the
MAX
button to toggle between Pt probe types. Press the
°C/°F
button to
confirm the Pt selection. Press the BACKLIGHT button to enter the reset
function. Press the
°C/°F
button to reset the calibration value to factory
settings. Press the
°C/°F
button to confirm the selection and resume
normal operation.