background image

3

G

ENERAL

M

ETER

S

PECIFICATIONS

1. 

DISPLAY

:  4-digit,  0.56"  (14.2  mm)  high  LED,  minus  sign  displayed  for

negative temperatures. 

Overrange/Underrange Input

: Flashing “

” or “

Overrange/Underrange Display

: “

   

” or “

   

2. 

POWER

: 85 to 250 VAC, 50/60 Hz, 6 VA

Isolation

: 2300 Vrms for 1 min. between input and supply (300 V working

voltage)

3. 

CONTROLS

: Three front panel push buttons for meter set-up. Rear terminal

input for disabling the front panel.

4. 

THERMOCOUPLE TYPES

: T, E, J, K, R, S, B, N or mV scale

5. 

RESOLUTION

: 1 degree for all types, or 0.1 degree for T, E, J, K and N only

6. 

THERMOCOUPLE RANGE AND ACCURACY

: All errors include NBS

conformity, cold junction effect and A/D conversion errors at 23°C after 60
minutes warm-up. Relative Humidity less than 85%.

7. 

INPUT IMPEDANCE

: 20 M

, all types

8. 

LEAD RESISTANCE EFFECT

: 20 µV/350 

Max Input Voltage Protection

: 70 VDC continuous

9. 

OPEN THERMOCOUPLE DETECTION

: Display Flashes: “

 

10. 

COLD JUNCTION COMPENSATION

: Automatic, 0.02 degree/degree.

Disabled for linear mV scale.

11. 

READING RATE

: 2.5 readings/second

12.

RESPONSE  TIME

: 2  seconds  to  settle  for  step  input  (increases  with

programmable digital filtering)

13. 

LOW FREQUENCY NOISE REJECTION

:

Normal  Mode  Rejection

: 45  dB  @  50/60  Hz  (may  be  improved  by

programmable digital filtering)

Common Mode Rejection

: 120 dB, DC to 50/60 Hz 

14. 

ENVIRONMENTAL CONDITIONS

:

Operating Temperature Range

: 0 to 50 °C

Storage Temperature Range

: -40 to 80 °C

Operating and Storage Humidity

: 85% max (non-condensing) from 0 to

50 °C

Span Drift

: 40 ppm/°C

Zero Drift

: 1 µV/°C

Altitude

: Up to 2000 meters.

15. 

CERTIFICATIONS AND COMPLIANCES

:

SAFETY

UL Recognized Component, File # E179259, UL3101-1, CSA C22.2 No. 1010-1

Recognized  to  U.S.  and  Canadian  requirements  under  the  Component
Recognition Program of Underwriters Laboratories, Inc. 

UL Listed, File # E137808, UL508, CSA C22.2 No. 14-M95

LISTED by Und. Lab. Inc. to U.S. and Canadian safety standards
Type 4X Enclosure rating (Face only), UL50

IECEE CB Scheme Test Certificate # UL/7470A/UL

CB Scheme Test Report # 03ME09282-08292003

Issued by Underwriters Laboratories, Inc.

IEC 1010-1, EN 61010-1: Safety requirements for electrical equipment for

measurement, control, and laboratory use, Part 1.

IP65 Enclosure rating (Face only), IEC 529

ELECTROMAGNETIC COMPATIBILITY

Emissions  and  Immunity  to  EN  61326:  Electrical  Equipment  for

Measurement, Control and Laboratory use.

Note:

1. Criterion A: Normal operation within specified limits.

16. 

CONSTRUCTION

: This unit is rated for NEMA 4X/IP65 indoor use. One

piece bezel/case. Flame resistant. Panel gasket and mounting clip included.

17. 

CONNECTIONS

: High compression cage-clamp terminal block

Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 4.5 inch-lbs (0.51 N-m) max.

18. 

WEIGHT

: 0.65 lbs. (0.24 Kg)

Class B

EN 55011

Emissions

Emissions:

0.5 cycle

Criterion A

EN 61000-4-11

Voltage dip/interruptions

3 V/rms 

Criterion A

EN 61000-4-6

RF conducted interference

1 kV signal

1 kV L-L, 

Criterion A

EN 61000-4-5

Surge 

2 kV signal

2 kV power

Criterion A

EN 61000-4-4

Fast transients (burst)

2 kV L&N-E power

10 V/m

Criterion A

EN 61000-4-3

Electromagnetic RF fields

8 kV air discharge

4 kV contact discharge

Criterion A

EN 61000-4-2

Electrostatic discharge

Immunity:

A

CCESSORIES

UNITS LABEL KIT (PAXLBK)

Each meter has a units indicator with backlighting that can be customized using

the Units Label Kit. The backlight is controlled in the programming.

Each meter is shipped with °F and °C overlay labels which can be installed into

the meter’s bezel display assembly.

TC TYPE

RANGE

ACCURACY

WIRE COLOR

T

-200 to +400ºC

-328 to +752ºF

0.8ºC

1.4ºF

blue

E

-200 to +1000ºC

-328 to +1832ºF

0.8ºC

1.4ºF

purple

J

-200 to +760ºC

-328 to +1400ºF

0.8ºC

1.4ºF

white

K

-200 to +1250ºC

-328 to +2282ºF

0.8ºC

1.4ºF

yellow

R

0 to +1768ºC

+32 to +3214ºF

2.1ºC

3.8ºF

black

S

0 to +1768ºC

+32 to +3214ºF

2.1ºC

3.8ºF

black

B

+150 to +1820ºC

+302 to +3308ºF

2.3ºC

4.1ºF

grey

N

-200 to +1300ºC

-328 to +2372ºF

0.8ºC

1.4ºF

orange

mV

0.01%

-10.00 to +80.00 mV

Содержание PAXLTC

Страница 1: ...the stability of the reading All set up data is stored in EEPROM which will hold data for a minimum of 10 years without power The meter has several built in diagnostic functions to alert operators of...

Страница 2: ...ns and Display 5 Programming the Meter 6 Calibrating the Meter 7 Troubleshooting 8 TABLE OF CONTENTS PAXL 0 0 TC Thermocouple Temperature Meter TC Accessories Part Numbers MODEL NO DESCRIPTION PART NU...

Страница 3: ...rds Type 4X Enclosure rating Face only UL50 IECEE CB Scheme Test Certificate UL 7470A UL CB Scheme Test Report 03ME09282 08292003 Issued by Underwriters Laboratories Inc IEC 1010 1 EN 61010 1 Safety r...

Страница 4: ...ING TC SENSOR Remove power and connect the negative thermocouple lead always red to TC Terminal 6 and the positive lead to TC Terminal 5 Be certain that connections are clean and tight If the thermoco...

Страница 5: ...Install them on Signal and Control cables as close to the unit as possible Loop the cable through the core several times or use multiple cores on each cable for additional protection Install line fil...

Страница 6: ...ution for the offset value is the same as the decimal point position programmed above The display offset is not available when mV indicator mode is selected for thermocouple type When the Input displa...

Страница 7: ...ype and verifying unit accuracy at various points over the range of the meter 10 to 80 mV 7 The factory settings for the programming parameters are shown in the previous section in the alternating dis...

Страница 8: ...wiring 1b Verify power IN DISPLAY 1 Program data error 1 Press PAR and check data set ups or IN DISPLAY 1 Input display out of range 2 Loss of data set ups DISPLAY WANDERS 1 Loss of data set ups JITTE...