SenSmart 3000 User's Manual
Document: UM-1003
Revision Level B
24
S E N S O R S E T T I N G S / I N F O
I n s t a l l S e n s o r
R e c a l l C a l H i s t o r y
I n p u t T y p e S MA R T
N e w S e n s o r 0 1/ 2 8 / 0 4
R e c e n t C a l 0 4 / 0 5 / 0 4
>
>
I N S T A L L N E W S E N S O R
E D I T t o i n s t a l l
N e w S e n s o r .
A n y o t h e r k e y
t o E x i t .
C A L I B R A T I O N D A T E S
R e c o r d 2 2 o f 2 2
C A L D a t e : 0 4 / 0 5 / 0 4
U P / D O WN t o s c r o l l
N E X T t o E X I T .
I N S T A L L N E W S E N S O R
H i s t o r i c a l C A L d a t a
Wi l l b e D e l e t e d .
E D I T t o A c c e p t
N E X T t o A b o r t
EDIT
EDIT
S e n s o r L i f e
>
S MA R T S E N S O R
T y p e : C a t - B e a d
S p a n : 10 0
Z e r o : 0
S N : x x x x x x
B o r n O n : 0 1/ 2 8 / 0 4
L a s t C a l : 0 4 / 0 5 / 0 4
A N Y k e y t o E x i t
Figure 5-8: Sensor Information Menus
Install New Sensor
should always be performed when a new
simple
sensor is installed. This
deletes historical CAL data and sets sensor life to 100% after initial calibration of the new
simple
sensor. The SenSmart 3000
Smart
sensor interface will automatically detect new smart sensors
and this menu is therefore not available with a smart sensor connected.
Recall Cal History
recalls each successful calibration. These dates may be reviewed by
scrolling with the UP / DOWN keys.
Input Type
indicates what kind of input or sensor the SenSmart 3000 is configured to accept and
is typically pre-configured at the factory. There are five Input Type possibilities consisting of
bridge, EC negative, EC positive, 4-20mA and Smart. Smart sensors upload sensor type and
other data to the SenSmart 3000 and may be viewed on the SMART SENSOR information
screen.
New Sensor
displays the date when a new sensor was last installed.
Recent Cal
displays the most recent calibration date.
5.6
CLOCK/DELAY SETUP:
Since the SenSmart 3000 is equipped with a Real Time Clock & Calendar
Time
and
Date
must
be set to correctly match its location. They are set at the factory in a 24 hour format but may
require adjustment to match the location’s time & date after shipment. Follow the procedure in
Configuration Using the Magnetic Wand
in section 3.3.
Warm Up
and
Cal Purge
time delays are also available to prevent unwanted alarm trips. Figure
5-9 shows the menu for these items.
C L OC K / D E L A Y S E T U P
T i me 12 : 3 5 : 4 2
D a t e 0 4 / 2 2 / 0 4
Wa r m U p S e c 12 0
C a l P u r g e S e c 10 0
Figure 5-9: Clock & Calendar / Delay Timer Menu