SenSmart 3000 User's Manual
Document: UM-1003
Revision Level B
20
L E L I N P U T 2
L C D C o n t r a s t A d j .
Mo d b u s R S 4 8 5 P o r t
S e c u r i t y
T e c h n i c i a n s ON L Y
S e n s o r T e mp
F i r mw a r e R e v .
>
>
>
>
>
“Sensor Temp” menu only present
with
“Arctic” smart sensor installed.
See section 7.1.
EDIT
L E L I N P U T 1
S y s t e m C o n f i g .
A l a r m S e t t i n g s
S e n s o r I n f o r ma t i o n
C l o c k / D e l a y s
D i a g n o s t i c s
H E L P
>
>
>
>
>
>
(Instructions)
Figure 5-2: Setup Menu Entry
5.3
System Configuration Menus:
The
System
Config
. group consists of two pages of menus as shown in Figure 5-
3. Each item’s
description follows in this section.
S Y S T E M C ON F I G. P G 1
Me a s u r e me n t N a me -
16
E u n i t s X X X X X X X X X X -
10
Z e r o ( 0 % )
(Eunits)
S p a n ( 10 0 % )
(Eunits)
D e c i ma l P o i n t s 0
C a l S p a n V a l 5 0
UP/
DOWN
S Y S T E M C ON F I G. P G 2
*Simple Sensor Installed **Smart Sensor Installed
>
>
>
*
“Restore” available only
without smart sensor installed.
**
“Upload” available only
with smart sensor installed.
R e a d o u t D e a d b a n d 0 %
T r a c k N e g a t i v e N O
T e mp C o mp e n s a t i o n
B a c k u p C o n f i g .
*R e s t o r e or **U p l o a d
C a l Ma r k e r 3 . 0 0 mA
T X S e n s o r L i f e N O
Figure 5-3: System Config. Menus
Measurement Name
may be edited to contain virtually any 16-character ASCII field. It is
typically used to describe the monitored point by user tag # or other familiar terminology.
Eunits
(engineering units) may have up to a 10 character ASCII field. Many common gases
have pre-configured Eunits based upon the sensor type and each may be edited in this menu as
described in
Configuration Using the Magnetic Wand
section 5-2.
Zero
(0%)
defines the reading to be displayed when 4mA (0%) is the SenSmart 3000 output.
Span
(100%)
defines the reading to be displayed when 20mA (100%) is the SenSmart 3000
output. The highest reading allowed is 9999 includes negative polarity sign and one decimal
point. Polarity is only indicated for negative readings.
Decimal Points
sets the resolution of the LCD readings and may be for 0, 1 or 2. Example:
ZERO readings for 0, 1 & 2 DP’s respectively are
0
,
0.0
&
0.00
.
Cal Span Value
sets what upscale value must be applied when performing Span calibrations.
Readout Deadband
allows forcing low values to continue to read zero. This is useful when
there are small amounts of background gases that cause fluctuating readouts above zero. The
highest amount of deadband allowed is 5%. The 4-20mA output is not affected by this menu
item.
Track Negative
,
set to NO, causes negative values to read the
Zero (0%)
value in data
displays. The CAL MODE readout displays negative values regardless of this setting and