November 5, 2010
Rev1.4
52
Drying agent
If the logger is closed up in a hotter, more humid environment than the deployment
environment and then plunged into cold water, internal condensation can result. Since
condensation may cause the circuitry to malfunction, it is advised in such circumstances
that, after the logger is set up and before it is closed for deployment, a small pack of
desiccant be placed on the battery side of the circuit board above the battery mount.
A desiccant plastic “pocket” should be installed before adding a desi-pack. The pocket is an
L-shaped piece of plastic designed to keep any moisture absorbed by the desi-pack off the
electronics.
If your application requires desiccant, please contact RBR for more information.
Repairs
We support all instruments and software that we manufacture. First line support is always
available by phone, fax or e-mail. Please contact us immediately if you are experiencing
problems with your RBR product. It is very important to contact us before returning your
instrument, as some difficulties can be easily solved on-site by the user. Please have the
model and serial number of the unit handy when you contact us.
There are no user-repairable parts of the logger. Any attempt at repair, whether successful
or not, without prior authorization from RBR Ltd. will void the warranty.
If it is necessary to return the product to RBR for an upgrade, repair, or calibration; please
review the detailed shipping information on our website before returning the unit.
Calibration
We recommend that you verify the calibration of your data logger before any critical deployment,
periodically once a year, or if you suspect the calibration to be out of specifications. Discuss
your calibration needs with RBR. In some cases, you will be recommended to return the
instrument to RBR to have it checked and re-calibrated. Please contact us for our current
calibration fees
TR-1060 loggers
1050 & 2050 series
XR series