background image

43

Appendix A - Unit functions

42

Appendix A - Unit functions

Lock

°

C/

°

F

Auto/All

Laserflash

Ltd. Access

Set Default

Backlight

Buzzer

TC-J/TC-K

NTC/TC

Time/Date

Photoflash

ON

ON

Open the battery compart-
ment and set the switches
ON or Off according to the
desired probe type.

NTC - thermistor

TC - thermocouple
Thermocouple type J
Thermocouple type K

Connect the probe to the input.
The input is located at the bot-
tom of the handle behind the
rubber cover.

Press MODE, until the desired
probe symbol (TC-K shown
here) appears. The probe
temperature is shown in the
lower part of the display. The
real time infrared temperature is
shown in the main display.

The Mode button - Thermocouple Settings

back to

Mode

button

The Emiss. button

To choose the emissivity of a mate-
rial, press EMISS.

The display shows a material
name, an emissivity value, and the
calculated temperature value. To
choose another material, use the
Up and Down keys.
Press ENTER  to activate this set-
ting.

To adjust the emissivity value,
press EMISS., when not in LOG
mode. Use the Up and Down keys
to select “Free“ (“Free” will have a
flashing underline).
Press EMISS again. “Free” is not
underlined, and the emissivity icon
flashes. Use the Up and Down keys
to adjust. Press ENTER (D) to acti-
vate this setting.

Содержание MX6

Страница 1: ...robat Reader choose the tab Book mark at the left to see the Table of Contents These headlines are linked with the corresponding chapters of the manual Just click on a headline to see the chapter in t...

Страница 2: ...with its basic functions It helps to get first results fast and easy It explains the software installation and how to use the unit with factory defaults Please take a bit time and follow up carefully...

Страница 3: ...1 port 15 MByte free diskspace on the hard drive The software works with Windows 98 ME XP and 2000 Note With a fresh set of batteries the unit can store photographs for up to 2 000 hours However we re...

Страница 4: ...synchronized with the temperature measurement True Spot laser Sighting Nameable locations Adjustable emissivity High Low alarm Data logger and a lot more Features Delivery Content If you see this ico...

Страница 5: ...Hardware Assistant will guide you through the installation process If you are asked where to look for the drivers choose CD ROM In most cases Windows will find the drivers automatically It could happe...

Страница 6: ...ure proceed with the following steps 1 Choose Setup in the software s menu bar 2 Go to Device Alarm and Display 3 Look for Date and Time and verify that the correct values are shown these values depen...

Страница 7: ...ion and look over it as shown below The embossed triangle will show you the approximate width of the photo The view finder guide Laser ON and OFF laser ON symbol The laser circle shows the spot size t...

Страница 8: ...ove the display The next location will be shown on the display Caution If you hear a longer beep and the LED above the display is shining red look at the display If you see Use Flash and a flash symbo...

Страница 9: ...ndow appears With the distance slider you could adjust the size of the laser circle drawing corresponding to the original size of the real laser circle NOTE This feature will only work properly if you...

Страница 10: ...tos and data A special file has been created automatically by the unit based on the factory defaults This file is called Logger and has the extension lgg Under special circumstances it may happen that...

Страница 11: ...of photos is 26 In the Low resolution mode you can take more than 100 photos The camera memory can store up to 106 photos This could be hel pful if you want to repeat a photo at a previous position Ne...

Страница 12: ...ta fields should be inserted in the photos after you have taken them All columns you choose here will be inclu ded in the final photo You can t delete any infor mation from the photo itself To open th...

Страница 13: ...ind out that something is outside the normal temperature range especially if the panel has live voltage If you monitor the panel over time you find out where the problems are before they become danger...

Страница 14: ...t will look as follows The routing graph Expert Mode The following dialog appears Now select in the curves menu up to five objects measured and set up their appearance in the graph Then choose the tim...

Страница 15: ...ble by clicking the button SAVE TO FILE In the subse quent window see next page enter the file name the file suffix mat is automatically appended Save the file in a directory on the drive of your choi...

Страница 16: ...shot should be taken at high alarm or low alarm The min interval s setting means that the unit only checks after this interval for the alarm val ues exceeded A photo will only be taken if an alarm is...

Страница 17: ...33 32 Expert Mode After taking the snapshot you can see the photos by clicking on the square markers in the graph...

Страница 18: ...ement of the unit Lock C F Auto All Laserflash Ltd Access Set Default Backlight Buzzer TC J TC K NTC TC Time Date Photoflash ON ON Sets the unit into con tinuous measuring mode Switches between degree...

Страница 19: ...has been stored at this logger position Camera is active at this logger posi tion Nothing has been stored Camera is inactive at this logger position Nothing has been stored The Display button The las...

Страница 20: ...unction can also be used to increase the accuracy for a narrow temperature range Change the time using the Up and Down keys Then press ENTER for each time segment to activate this time setting The seg...

Страница 21: ...re values during a measurement at the bottom of the display press MODE until the two values appear To activate the MIN mode press MODE until the MIN symbol appears The measured minimum temperature is...

Страница 22: ...be temperature is shown in the lower part of the display The real time infrared temperature is shown in the main display The Mode button Thermocouple Settings back to Mode button The Emiss button To c...

Страница 23: ...wn keys to select the material name Free which will be shown in the display Press the emissivity button again until the emissivity sign flashes Use the arrow keys to change the emissivity value until...

Страница 24: ...in the far field 33ft 10m is 35 1 CLOSE FOCUS MODEL Optical Chart 500 20 1 0 0 0 93 25 23 SPOT DIA IN SPOT DIA mm DISTANCE SENSOR TO OBJECT mm DISTANCE SENSOR TO OBJECT IN FOCUS POINT D S 50 1 FAR FI...

Страница 25: ...r 0 93 Wood 0 94 oxidized opaque over 20 mils natural Table of emissivity values of common materials The DIP switches are located in the battery compartment 8 In the beginning we recom mend you use th...

Страница 26: ...Data Logger Sometimes a long beep and the red LED above the display shines to indicate a problem check your display to find out which problem has occured Symptom Display Code Problem Action Photo qua...

Страница 27: ...camel s hair brush Wipe the surface with a moist cotton swab The swab may be moistened with water or a water based glass cleaner NOTE DO NOT use solvents to clean the plas tic lens Cleaning the Hous...

Страница 28: ...curate temperature readings Thermal Shock This instrument conforms to the following standards EMC EN 61326 1 Safety EN 61010 1 1993 A2 1995 EN 60825 1 1994 This product herewith complies with the requ...

Страница 29: ...ver is greater at 25 C 77 F 25 C 45 F Optical Resolution 60 1 19mm spot size at 1 15m Standard Focus 0 75in spot size at 3 8 feet Optical Resolution 50 1 6mm spot size at 0 3m Close Focus 0 24in spot...

Страница 30: ...he printed manual Caution Wait for unit Auto Shut down before battery change You will see the message Shut down for a few seconds max 30 sec on the display after the unit is used in LOG mode Download...