Appendix
36
MP150 Protocol
Description
Code
Para-
meter
Explanation
Factory
default
Firmware
Revision
Hold Time
HT
ndddd
Hold time in Hold Mode 1 with:
n=1..3: sector/interface
dddd: time in tenth of a second
0
Hold Value
HV
nc
n=1..3: sector/interface c=’P’: Hold peak
value
c=’V’: Hold valley value
P
Internal
Temperature
I
Returns
float
Returns the internal temperature in °C
-
IP address
IP
d.d.d.d
d = 0 ... 255
192.168.42.30
Combination of IP,
NM and PO
IP_NM_PO
d.d.d.d
n.n.n.n
port
d = 0 ... 255
n = 0 ... 255
port = 1025 ... 65534
192.168.42.30
255.255.255.0
2727
Keep alive time of
the TCP/IP-
connection
KeepAlive
time intvl
probes
time = tcp_keepalive_time
intvl = tcp_keepalive_intvl
probes = tcp_keepalive_probes
(see section 2.1.1
240 10 6
Line Count
LC
ddd
Count of lines to send after STX
ddd = 1 … 768
1
Line Mode
LM
hexa
LM = 0: no Appendix to Line Transfer
LM = 1: Sector Values
LM = 2: Zone Values
LM = 5: Sector Values (like LM =1) with
bit 15 = Sector Alarm,
bit 14 = serial Alarm
LM = 6: Zone Values (like LM =2) with
bit 15 = Zone Alarm,
bit 14 = serial Alarm
LM = 8 … F
doubles line modes 0 … 6
but provides additional framing
LM = 11: the six bytes of the traditional
sector values are changed in their
contents:
previous current output 1: internal
temperature in hundredth of °C (MSB
first!),
previous current output 2: open for future
use,
previous current output 3: open for future
use.
1
MirroR lines
MR
d
mirror (or rotate by 180°) each line
d = 0: don’t mirror
d = 1: mirror on
0
3.41
Net Mask
NM
d.d.d.d
d=0-255
255.255.255.0
Parameter Load
PL
-
Loading of all parameters from Flash-ROM done at reset
Port for the
Ethernet interface
PO
d
d = 1025 ... 65534
2727
Parameter Store
PS
-
Storing of all user parameters to Flash-
ROM
-