Dedicated Commands
MP150
Protocol
17
5
Dedicated Commands
Additional description to certain commands is given in the following section. A summary of all
commands is to be found in a summarizing list of section
5.1
Start-up Parameter
While the scanner is in operation, all parameters are hold in R
AM
. Various protocol commands effect
changes of these R
AM
based parameters.
When powering on the scanner, parameters are transferred from Flash-R
OM
into R
AM
. Unless the
PS
command is issued, powering off and on the scanner will result in the original parameter values being
loaded into R
AM
(the loss of the changed parameters).
The
PS
command stores the changed parameters from the R
AM
into the Flash-R
OM
. By moving the
changed parameters to this permanent memory, the parameters become the scanners start-up
parameters.