Ratek OM8 Orbital Shaker – User Manual 

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

 

 

----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- 

Page 11 of 15

 

 

 

Storing & Relocating The Orbital Shaker

 

 

The Orbital Shaker should be stored out of direct sunlight at an ambient temperature below 30° Celsius in a clean and dry 
location which meets the environmental conditions required as detailed in the technical specifications of this User Manual. 
 

-

 

Switch the power to “Off” (element 1 in Figure 1) 

 

-

 

Unplug the equipment from the mains power supply outlet. 

 

-

 

Ensure the Orbital Shaker is clean and dry. If required, clean it according to the section of this User Manual titled “Routine 
Cleaning & Maintenance”. 

 

-

 

Relocate it to an appropriate storage location. Ensure the Orbital Shaker is stored in a clean and dry location away from 
potential damage by accidental knocks and bumps. 

 

Optional Accessories 
 

The OM8 platform has a Nitrile rubber mat supplied as standard and will be sufficient to hold most flasks and containers at low 
speeds. 
 
For higher speeds a suitable retaining system would be required.  The following optional clips and racks are available from your 
Ratek dealer. 

 

 

 

 

RR25 Universal Rack 

 
 

This rack consists of slotted channels attached to the mixers platform, and 6 stainless steel rubber covered bars.  The bars 
can be moved horizontally enabling flasks, test tube racks and most other types of containers to be secured to the platform. 

 

 

 

TCT25 Tulip Clip Tray

 

 

 

 

 

 

This tray is suitable for holding tulip clips dedicated to conical flasks.  Tulip clips are available to suit flasks from 25ml to 
2000ml.   
 
The tray consists of a stainless steel platform with 4mm threaded inserts, and is made to fit within the OM8 platform. 
 
The following Tulip clips are available: 
 

TC25

 to suit 25ml flasks                                            

TC500

 to suit 500ml flasks 

TC50

 to suit 50ml flasks                                            

TC1000

 to suit 1000ml flasks 

TC100

 to suit 100ml flasks                                        

TC2000

 to suit 2000ml flasks 

TC250

 to suit 250ml flasks 

 
To avoid damage to the Tulip clips, care should be taken to use the correct size clip for the flask to be shaken (eg: TC500 for 
500ml conical flask).  Continual over bending of the spring sides to accommodate incorrect flask sizes will cause them to 
fracture. 

 

 

 

Routine Cleaning And Maintenance 

 
To maintain the Orbital Shaker in good, safe working order and ensure maximum product lifespan, regular cleaning and general maintenance is 
required. The Orbital Shaker should be cleaned at least once every month for a unit being used on a daily basis, for infrequently used Orbital Shakers 
a cleaning frequency of once every 3 months is recommended. On each occasion, the general maintenance routine should be employed following 
cleaning. 
  
 

Содержание OM8

Страница 1: ...015 English www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Australia User Manual printed in Australia...

Страница 2: ...the product to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product into the supplied...

Страница 3: ...TY CONSIDERATIONS 5 5 5 5 Conditions of Operation 5 Safety Labels And Markings 6 OPERATING PROCEDURES 6 6 6 6 Identification of Controls Functions 6 Safety Warnings 7 Safety Recommendations 7 Preparat...

Страница 4: ...r the Operator Any item that is not suitably secured to the mixing platform Suitable Environments For The Orbital Shaker The Orbital Shaker is intended for use in a clean laboratory environment only w...

Страница 5: ...ng in how to perform their work safely in accordance with any local operational health and safety regulations The operator and all users should be well versed in local emergency procedures as per the...

Страница 6: ...as a rotating platform which should not be touched whilst it is moving Keep all body parts away from the moving platform whilst in operation Always stop the platform using the front panel speed contro...

Страница 7: ...kplace administrator or electrical maintenance staff to determine if a protective earth power socket is available A surge protected power outlet is strongly recommended as it provides some protection...

Страница 8: ...inimum unobstructed distance of 500 millimetres in front of the Orbital Shaker to allow adequate room for the user to maintain a safe operating distance of 300 millimetres CAUTION The Orbital Shaker i...

Страница 9: ...erly rated protective earthed wall mounted power supply outlet If there are double adapters or oversized DC power packs causing obstruction of the mains power lead plug these should first be removed E...

Страница 10: ...weight The counterweight is located beneath the platform carrier It is adjustable and has 4 positions The counterweight should be set for the expected load to enable the shaker to operate smoothly wit...

Страница 11: ...overed bars The bars can be moved horizontally enabling flasks test tube racks and most other types of containers to be secured to the platform TCT25 Tulip Clip Tray This tray is suitable for holding...

Страница 12: ...d receptacles are completely dry before restoring electrical power Unplug the equipment from the mains power supply outlet Allow the Orbital Shaker to air dry or use a dry cloth to remove water from a...

Страница 13: ...etres Temperature 0 Celsius to 45 Celsius Maximum relative humidity 80 for temperatures up to 31 Celsius decreasing linearly to 50 relative humidity at 40 Celsius MAINS supply voltage fluctuations up...

Страница 14: ...a 3 year parts and 12 months labour return to base warranty effective from the date of purchase The product is manufactured in Melbourne Australia The warranty is offered by Ratek Instruments Pty Ltd...

Страница 15: ...refund Ratek Instruments will pay your reasonable direct expenses of claiming under this Warranty You may submit details and proof of your expense claim to Ratek Instruments for consideration This Wa...

Отзывы: