Command Reference
R&S
®
ZNC
697
User Manual 1173.9557.02 ─ 13
Calibration type
Parameter in
[SENSe<Ch>:]CORRec-
tion:COLLect:METHod
Available error terms (depending on port
numbers)
TOSM
TOSM
'SCORR1' to 'SCORR12'
(at present the isolation terms 'SCORR4' and
'SCORR10' are not included)
TOM, TSM, TRM, TRL,
TNA
TOM | TRM | TRL | TNA
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK' (for reading
and writing)
'G11' ... 'G22' and 'H11', 'H12', 'H22' (for reading
only; the 'H21' matrix elements are normalized
to 1)
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude
<OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Tip:
Use the
[SENSe<Ch>:]CORRection:LOSS<PhyPt>...
commands to define the
complete set of loss offset parameters.
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude
is equivalent to
.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-independent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.0001 dB
*RST:
0 dB
Default unit: dB
[SENSe<Ch>:]FREQuency:MODE
<FreqSweep>
Selects the sweep type and defines which set of commands controls the stimulus fre-
quency.
Tip:
The command
provides a complete list of sweep
types.
Suffix:
<Ch>
.
Channel number
Parameters:
<FreqSweep>
CW | FIXed | SWEep | SEGMent
Linear or logarithmic frequency sweep, depending on the selected
spacing (
LINear |
LOGarithmic
). The frequency range is set via
R&S ZVR/ZVAB Compatible Commands