Command Reference
R&S
®
ZNC
442
User Manual 1173.9557.02 ─ 13
Example:
*RST; SWE:TYPE POW
Select a power sweep with default CW frequency and sweep
range.
CALC:STAT:NLIN:COMP:LEV 2
Define a compression value of 2 dB.
CALC:STAT:NLIN:COMP:RES?
Query the compression point results <Cmp In>, <Cmp Out>. An
execution error message (error no. -200) is returned if no com-
pression point is found.
CALC:STAT:NLIN:COMP ON
Display the compression point result in the diagram area.
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
<Boolean>
Displays or hides the compression point result in the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - statistical info field on or off.
*RST:
OFF
Example:
See
Manual operation:
See
CALCulate<Chn>:STATistics:RESult?
<Result>
Returns a single statistical parameter of the trace no.
<Chn>
or all parameters. It is not
necessary to display the info field (
ON
)
before using this command.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference