CALCulate<Chn>:STATistics:EPDelay[:STATe]
<Boolean>
CALCulate<Chn>:STATistics:MMPTpeak[:STATe]
<Boolean>
CALCulate<Chn>:STATistics:MSTDdev[:STATe]
<Boolean>
These commands display or hide the "Phase/El Length" results, the "Min/Max/Peak-
Peak" results, and the "Mean/Std Dev" results in the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - statistical info field on or off.
*RST:
OFF
Example:
See
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel
<Stop>
Defines the compression value x for the compression point measurement (
).
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Stop>
Stop value of the evaluation range.
Range:
+0.01 dBm to +100.00 dBm
*RST:
1 dBm
Default unit: dB
Example:
See
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
Returns the x-dB compression point of an S-parameter or ratio measured in a power
sweep. The compression value x is set via
.
The response contains two numeric values:
●
<Cmp In>
–
stimulus level at the compression point in dBm.
●
<Cmp Out>
–
sum of <Cmp In> plus the magnitude of the measured response value
at the compression point in dBm.
Suffix:
<Chn>
.
Channel number used to identify the active trace.