GUI Reference
R&S
®
ZNB/ZNBT
506
User Manual 1173.9163.02 ─ 55
On "Apply", the raw calibration data are transferred to METAS VNA Tools, which take
over the calculation of measurement results and uncertainties:
is
enabled automatically.
In parallel to the "METAS Calibration", a regular VNA calibration without uncertainties
is created. This calibration can be activated if "METAS Cal Active" is unchecked.
6.11.5.3
METAS Verification Dialog
This dialog allows you to set up and perform a
.
For background information see
Access:
CHANNEL – [CAL] > "METAS Cal" > "Start Verification"
Step 1: Verification Kit
Allows you to select the verification kit to be used. All other settings are taken from the
active
Verification Kit
Allows you to specify the verification kits to be used for the related ports.
Verification standards are characterized individually, including uncertainties. So initially
no verification kits are defined. Use
kit.
Import Verification Kit...
Opens the "Import Verification Standards with Uncertainties" dialog.
Cal Softtool