2.10.4 Configuration for manual remote operation
The R&S UPV can be operated manually from an external computer via a network con-
nection. The Windows program Remote Desktop Connection is used for operation, see
chapter 4.16, "Manual Remote Operation"
Manual remote control of the audio analyzer is possible under the following conditions:
●
the Windows 95 operating system or higher and the Remote Desktop Connection
program are installed on the external computer and a LAN interface is configured
for the network
●
the audio analyzer and the computer are connected via LAN (see
"Configuration of R&S UPV for Network Operation"
●
the Remote Desktop Connection program is activated on the audio analyzer (see
chapter 2.10.4.1, "Activating the Remote Desktop Connection program at the R&S
UPV"
●
on the external computer, the R&S UPV data have been entered in the Remote
Desktop Connection program (IP address or computer name of R&S UPV in the
network (see
chapter 2.10.2.3, "Querying the computer name"
●
the login at the external computer for R&S UPV was performed with the correct
user name (instrument) and the correct password (instrument) (see
ter 2.10.4.2, "Starting manual remote operation at the external computer"
2.10.4.1
Activating the Remote Desktop Connection program at the R&S UPV
Unauthorized access possible to audio analyzer
After activating Remote Desktop, access to this R&S UPV is possible for any user in
the network who knows the computer name and the login data of the audio analyzer.
Activating Remote Desktop
1. "Start – Settings – Control Panel": Click on "Performance and Maintenance".
2. "Click on System" and select "Remote" in the tab menu.
3. Activate the "Allow users to connect remotely to this computer" check box.