Signal Generation Control
Satellite Navigation
202
User Manual 1178.6379.02 ─ 01
18.3 Clock Settings
Access:
► Select "GNSS > Clock".
This tab provides access to the settings necessary to select and configure the
clock signal, like the clock source and clock mode.
This section focuses on the available settings.
For information on how these settings affect the signal, refer to section "Basics on ..."
in the R&S
SMW user manual.
Defining the Clock
The provided clock signals are not dedicated to a particular connector. They can be
mapped to one or more globally shared USER and the local T/M/C connectors.
Use the
Local and Global Connector Settings
to configure the signal mapping, the
polarity, the trigger threshold, and the input impedance of the input connectors.
To route and enable a trigger signal, perform the following
general steps
:
●
Define the signal source, that is select the "Clock > Source".
●
Define the connector (USER or T/M) where the selected signal is provided.
Use the
Local and Global Connector Settings
Settings:
............................................................................................ 203
Clock Settings