Remote Commands Reference
R&S
®
RTC1000
293
User Manual 1335.7352.02 ─ 02
TRIGger:A:UART:MODE
<Mode>
Specifies the trigger mode for UART/RS-232 interfaces.
Parameters:
<Mode>
BSTart | SBIT | NTHSymbol | SYMBol | PATTern | PERRor |
FERRor | BREak
BSTart
Burst start. Sets the trigger to the begin of a data frame. The
frame start is the first start bit after the idle time.
SBIT
Start bit. The start bit is the first low bit after a stop bit.
NTHSymbol
Sets the trigger to the n-th symbol of a burst.
SYMBol
Triggers if a pattern occurs in a symbol at any position in a burst.
PATTern
Triggers on a serial pattern at a defined position in the burst.
To define the pattern, use
and
PRERror
Parity Error: Triggers if a bit error occured in transmission.
FERRor
Triggers on frame error.
BREak
Triggers if a start bit is not followed by a stop bit within a defined
time. During the break the stop bits are at low state.
*RST:
SBIT
TRIGger:A:UART:PATTern
<DataPattern>
Defines the bit pattern as trigger condition.
Parameters:
<DataPattern>
Binary pattern with max. 32 bit. Characters 0, 1, and X are
allowed.
*RST:
1 = "00000001"
TRIGger:A:UART:PLENgth
<PatternLength>
Defines how many symbols build up the serial pattern.
Protocol Analysis
Содержание RTC1000
Страница 1: ...R S RTC1000 Digital Oscilloscope User Manual User Manual 1335 7352 02 02 S 2...
Страница 57: ...Trigger R S RTC1000 57 User Manual 1335 7352 02 02 Figure 5 3 External trigger signal External Trigger Input...
Страница 225: ...Remote Commands Reference R S RTC1000 225 User Manual 1335 7352 02 02 Suffix m 1 6 Selects the measurement Measurements...