Remote Commands Reference
R&S
®
RTC1000
258
User Manual 1335.7352.02 ─ 02
DVM<m>:RESult[:ACTual]?
Returns the current value of the indicated measurement.
Return values:
<CurrentValue>
Numeric value
Example:
DVM2:SOUR CH2
DVM2:TYPE DCRM
DVM2:RES?
<-- 7.089E-01
A RMS measurement is performed on measurement place 2, on
channel 2. The result is 708,9
mV.
Usage:
Query only
DVM<m>:RESult[:ACTual]:STATus?
Returns the result value and the status of the result.
The status is the decimal representation of a 4-bit register value:
●
Bit 0 = 1: result is valid
●
Bit 1 = 1: no result available
●
Bit 2 = 1: clipping occurs
●
Bit 3 = 1: no period found
Return values:
<ResultAndStatus>
<Value>,Status
Example:
DVM:SOUR CH1
DVM:TYPE MEAN
DVM:RES:STAT?
<-- 4.968E-01,5
The result value of the mean measurement on channel 1 is
496.1
mV. The result status is 5 (decimal) = 0101 (binary). That
means, the result is valid (bit 0 = 1), and the signal is clipped by
the limits of the ADC range (bit 3 = 1).
Usage:
Query only
DVM<m>:RESult:RESet
Resets the digital voltmeter function.
Suffix:
<m>
.
1..2
Selects the channel resp. the secondary measurement type.
Usage:
Event
Digital Voltmeter