Remote Control Commands
R&S
®
NRX
243
User Manual 1178.5566.02 ─ 05
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MEASurement:PULSe:SEParation[:STATe]
<value>
Enables or disables the display of the gap between two pulses.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<value>
*RST:
OFF
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MEASurement:RRELation
<refRelation>
Selects how the threshold parameters are to be interpreted, either voltage related or
power related.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<refRelation>
POWer | VOLTage
*RST:
POWer
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MEASurement:SELection
<traceNo>
Selects the displayed trace.
Suffix:
<Window>
.
1 to 4
Measurement channel
Parameters:
<traceNo>
Range:
1 to 100
*RST:
1
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:NEGative:
DURation[:STATe]
<value>
Enables or disables the display of the fall time of the first detected pulse. The fall time
is the time the signal requires to change from high to low level.
Suffix:
<Window>
.
1 to 4
Measurement channel
Measurement Settings and Results