Measurement Types and Result Displays
R&S
®
NRX
76
User Manual 1178.5566.02 ─ 05
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:PULSe:PERiod[:STATe]
Duty Cycle ← Time tab
Pulse period
Duty cycle =
Pulse width
The ratio is expressed as a value between 0 and 1.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:PULSe:DCYCle[:STATe]
Pulse Off Time ← Time tab
Time between the first negative edge and the subsequent positive edge of the pulse in
seconds, where the edges occur at crossings of the mid threshold. During this time, the
pulse remains at the pulse base level.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:PULSe:SEParation[:
STATe]
Rise Time ← Time tab
Time the pulse requires to transition from the pulse base level to the pulse top level.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:POSitive:
DURation[:STATe]
Fall Time ← Time tab
Time the pulse requires to transition from the pulse top level to the pulse base level.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:NEGative:
DURation[:STATe]
Start Time ← Time tab
Time when the signal passes through the medial reference power level with rising
edge, referenced to the delayed trigger event. Indicates the start point of the first power
pulse within the analysis window.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence[:STATe]
Stop Time ← Time tab
Time when the signal passes through the medial reference power level with falling
edge, referenced to the delayed trigger event. Indicates the stop point of the first power
pulse within the analysis window.
Pulse Analysis