Measurement Types and Result Displays
R&S
®
NRX
79
User Manual 1178.5566.02 ─ 05
Low Ref. ← Power tab
.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:POWer:PULSe:
LREFerence[:STATe]
Configuration tab
Defines the reference levels for the pulse timing. All values are specified in percent of
the pulse amplitude. The settings on this tab are trace-specific.
Reference Levels relate to ← Configuration tab
Selects whether the reference levels are voltage-related or power-related.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:RRELation
Algorithm ← Configuration tab
Selects the analysis algorithm for detecting the pulse top and the pulse base power of
a pulsed signal. From these power levels, the reference levels are derived.
"Histogram"
Analyzes the histogram of the trace data. The average of all points
representing the pulse top is taken as pulse top power. Similarly, the
pulse base is determined.
This algorithm is suitable for most pulse signals.
"Integration"
Fits a rectangle pulse of the same energy into the pulse signal as a
reference and thus determines the pulse top power.
This algorithm is recommended for:
●
Pulse signals with modulation
●
If the pulse energy is considered
For example, if you compare the measurement result to a mea-
surement result of a thermal power sensor.
Pulse Analysis