Remote control commands
R&S
®
NRPxxA(N)
95
User Manual 1177.6017.02 ─ 08
Example:
IGAM:PHAS?
Query
-1.02
Response
Usage:
Query only
[SENSe<Sensor>:]IGAMma:EUNCertainty?
Queries the expanded (k = 2) uncertainty of the magnitude of the complex input reflec-
tion coefficient
Γ
in
. Following gamma correction, the residual mismatch uncertainty
becomes so small that it is practically negligible.
Example:
IGAM:EUNC?
Query
5.000000E-03
Response
Usage:
Query only
9.8.3.6
Using the S-Parameters program
The S-Parameters program helps loading an S-parameter table into the power sensor.
The S-Parameters program is part of the R&S
NRP Toolkit, see
To start the S-Parameters program
► In the Windows start menu, select "NRP Toolkit" > "S-Parameter Update Multi".
Configuring basic measurement parameters