![R&S NRP Series Скачать руководство пользователя страница 120](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882120.webp)
Remote control commands
R&S
®
NRPxxA(N)
120
User Manual 1177.6017.02 ─ 08
Parameters:
<gateway>
Example:
SYST:COMM:NETW:IPAD:GAT '192.168.10.254'
Sets
192.168.10.254
as IP address of the default gateway.
Manual operation:
See
SYSTem:COMMunicate:NETWork:IPADdress:INFO?
Effective only for the R&S
NRP LAN power sensors.
Queries the network status information.
Usage:
Query only
SYSTem:COMMunicate:NETWork:IPADdress:MODE
<mode>
Effective only for the R&S
NRP LAN power sensors.
Sets whether the IP address is assigned automatically or manually.
Parameters:
<mode>
AUTO | STATic
AUTO
Assigns the IP address automatically, provided the network sup-
ports DHCP.
STATic
Enables assigning the IP address manually.
*RST:
AUTO
Example:
SYST:COMM:NETW:IPAD:MODE AUTO
The IP address is assigned automatically.
Manual operation:
See
SYSTem:COMMunicate:NETWork:IPADdress:SUBNet:MASK
<netmask>
Effective only:
●
For the R&S
NRP LAN power sensors.
●
SYSTem:COMMunicate:NETWork:IPADdress:MODE
is set to
STATic
.
Sets the subnet mask.
Parameters:
<netmask>
Example:
SYST:COMM:NETW:IPAD:SUBN:MASK '255.255.255.0'
Sets
255.255.255.0
as subnet mask.
Manual operation:
See
Configuring the system