![R&S NRP Series Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882113.webp)
Remote control commands
R&S
®
NRPxxA(N)
113
User Manual 1177.6017.02 ─ 08
SYSTem:INITialize
Sets the power sensor to the standard state, i.e. the default settings for all test parame-
ters are loaded in the same way as with
. The power sensor then outputs a com-
plete list of all supported commands and parameters. With the command, the remote-
control software can automatically adapt to the features of different types of sensors
with different functionality.
Usage:
Event
SYSTem:LANGuage
<language>
Selects an emulation of a different command set.
Parameters:
<language>
SCPI
*RST:
SCPI
SYSTem:LED:COLor
<color>
Effective if the operating mode of the LED is set to user-defined,
USER
.
Sets the color and the flash code of the system status LED.
Parameters:
<color>
Hexadecimal code described as
0x0krrggbb
with
k
= 0: steady on; k
= 1: slowly flashing; k
= 2: fast flashing
rr
= red
gg
= green
bb
= blue
In NRP legacy communication, the parameter is a standard deci-
mal number, representing the corresponding hexadecimal code.
Range:
0x00 to 0x02FFFFFF
*RST:
0x00A0A0A0
Example:
SYST:LED:MODE USER
Sets the system status LED operating mode to user.
SYST:LED:COL #H01a00000
The LED flashes slowly in red.
SYSTem:LED:MODE SENSor
Sets the system status LED operating mode back to the sensor
internal settings.