![R&S NRP Series Скачать руководство пользователя страница 146](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882146.webp)
Remote control basics
R&S
®
NRPxxA(N)
146
User Manual 1177.6017.02 ─ 08
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Lower limit fail
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 11-12: Operation lower limit fail status registers
Querying the register:
●
STATus:OPERation:LLFail:CONDition?
●
STATus:OPERation:LLFail[:SUMMary][:EVENt]?
Table 11-14: Used operation lower limit fail status bits and their meaning
Bit no.
Short description
Bit is set if
1
Lower limit fail
Measured value is below the lower limit value.
11.2.5.6
Operation upper limit fail status register
The
CONDition
registers contain information whether a measured value currently
exceeds a configured upper limit. The
EVENt
registers contain information whether a
measured value exceeded an upper limit value since the last query.