Operating Manual 1302.6163.12 - 03
4.127
R&S ESU
Instrument Functions
Analyzer Mode
When quitting the submenu, the program will return to the frequency domain pro-
vided it was active before. The original span is restored so the desired measurement
can now be performed with the accurately set gate.
Remote command:
--
GATE MODE
LEVEL | EDGE
The
GATE MODE LEVEL/EDGE
softkey selects the trigger mode. Gated sweep is
possible in the level-triggered as well as in the edge-triggered mode.
If level triggering is selected, the
GATE LENGTH
softkey becomes inactive and can-
not be operated.
Remote command:
SWE:EGAT:TYPE EDGE
POLARITY
POS | NEG
The
POLARITY POS/NEG
softkey controls the polarity of the
EXT TRIGGER/GATE
control line.
In case of level triggering the sweep is stopped by
POLARITY POS
and a logic ’0’
signal; the signal ’1’ will restart the sweep after the
GATE DELAY
time has elapsed.
In case of edge triggering the sweep is continued on a ´0´ to ´1´ transition for the
duration of
GATE LENGTH
after a delay (
GATE DELAY)
has elapsed.
Changing the polarity automatically implies a transition of the trigger-edge polarity
(
POLARITY
softkey in the higher menu).
Remote command:
SWE:EGAT:POL POS
GATE DELAY
The
GATE DELAY
softkey activates the window for setting the delay time between
the gate signal and the continuation of the sweep.
This may be useful for taking into account a delay between the gate signal and the
stabilization of an RF carrier for example.
As gate delay, values between 125 ns and 100 s may be set. The position of the
delay on the time axis in relation to the sweep is indicated by the line labelled
GD
.
As there is a common input signal for trigger and gate if
EXTERN
or
IF POWER
is
selected, changes to the gate delay will affect the trigger delay (
TRIGGER OFFSET
)
as well.
Remote command:
SWE:EGAT:HOLD 1US
GATE LENGTH
The
GATE LENGTH
softkey activates the window for setting the sweep duration of
R&S ESU in the edge-triggered mode.
Values between 125 ns and 100 s may be set for the gate length. The length of the
gate in relation to the sweep is indicated by the line labelled
GL.
This softkey is only available if
GATE MODE EDGE
(edge triggering) has been
selected.
Remote command:
SWE:EGAT:LENG 100US
SWEEPTIME
The
SWEEPTIME
softkey enables the user to change the time axis to obtain a
higher resolution for positioning gate delay and gate length.
When this is to be done, the sweep time temporarily changes; the original value is
restored when the menu is quit.
Remote command:
--
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...