Remote Control Commands
R&S
®
NRX
424
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]BURSt:MODE
<mode>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<mode>
AUTO | USER
*RST:
AUTO
[SENSe<Sensor>:]BURSt:PERiod
<value>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:PERiod
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.1
Default unit: s
[SENSe<Sensor>:]BURSt:WIDTh
<width>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:WIDTh
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<width>
Range:
0.0 to 1.0
*RST:
0.01
Default unit: s
[SENSe<Sensor>:]BWIDth:VIDeo
<mode>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:VBWidth:ENUM
R&S NRP2 Compatibility
www.allice.de
Allice Messtechnik GmbH